微型种植体长度对骨界面应力分布的影响 被引量:6

The effect of mini-implant lengths on stress distributions in peri-implant surface
下载PDF
导出
摘要 目的探讨微型种植体长度对骨组织内应力分布的影响。方法建立直径1.6 mm,长度分别为6、8、10、12 mm的种植体-下颌骨三维有限元模型,垂直植入种植体,给以1.96 N的水平向前及前上方向的载荷,记录并分析不同情况下的应力分布。结果种植体水平向前载荷的应力峰值范围为3.500~3.765 MPa,位移峰值范围为1.266~1.288μm;前上方向载荷的应力峰值范围为4.075~4.510 MPa,位移峰值范围为1.668~1.694μm。水平向前载荷的应力峰值及位移峰值均低于前上方向载荷的峰值。结论在所研究的尺寸(6~12 mm)范围内,种植体长度对骨界面应力影响不大,而种植体载荷方式对骨界面应力分布有重要影响。 Objective To observe the effect of mini-implant lengths on stress distribution in peri-implant surface.Methods The 3D finite element analysis mandible and mini-implant models with diameter of 1.6 mm,lengths of 6,8,10 and 12 mm were established.The mini-implants were inserted into designed site of mandibular vertically,respectivly.A force of 1.96 N were applied mesioly and 45° tilted mesio-vertically in models.The stress distribution under every condition was recorded and analyzed.Results When load was applied mesially,the maximum stress varied from 3.500 to 3.765 MPa,the maximum displacement varied from 1.266 to 1.288 μm.When load was applied 45°tilted mesio-vertically,the maximum stress varied from 4.075 to 4.510 MPa,the maximum displacement varied from 1.668 to 1.694 μm.All of the maximum stress and displacement of loading mesially were lower than loading mesio-vertically.Conclusion The change of the mini-implant length within 6-12 mm don't show much influence on the stress distribution.The loading type is an important factor influencing stress and displacement of peri-implant bone.
出处 《华西口腔医学杂志》 CAS CSCD 北大核心 2011年第1期27-30,共4页 West China Journal of Stomatology
基金 河北省科技支撑计划基金资助项目(08276101D-4)
关键词 微型种植体 应力分布 三维有限元分析 mini-implant stress distribution finite element analysis
  • 相关文献

参考文献12

  • 1Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage[J]. Am J Orthod Dentofacial Orthop, 2006, 130(1): 18-25. 被引量:1
  • 2单丽华,董福生,宫伟伟,张雄.微型种植体植入角度对骨界面应力峰值的影响[J].中国组织工程研究与临床康复,2008,12(52):10227-10230. 被引量:5
  • 3Lim SA, Cha JY, Hwang CJ. Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length [J]. Angle Orthod, 2008, 78(2):234-240. 被引量:1
  • 4Block MS, Delgado A, Fontenot MG. The effect of diameter and length of hydroxylapatite-coated dental implants on ultimate pul- lout force in dog alveolar bone[J]. J Oral Maxillofac Surg, 1990, 48(2) : 174-178. 被引量:1
  • 5Freire JN, Silva NR, Gil JN, et al. Histomorphologic and histo- mophometric evaluation of immediately and early loaded mini- implants for orthodontic anchorage[J]. Am J Orthod Dentofacial Orthop, 2007, 131 (6) : 704.e1-e9. 被引量:1
  • 6De Panw GA, Dermaut LR, Johansson CB, et al. A histomor- phometric analysis of heavily loaded and non-loaded implants[J]. Int J Oral Maxillofac Implants, 2002, 17(3):405-412. 被引量:1
  • 7Himmlova L, Dostalova T, Kacovsky, A, et al. Influence of im- plant length and diameter on stress distribution: A finite ele- ment analysis[J]. J Prosthet Dent, 2004, 91 (1):20-25. 被引量:1
  • 8Rieger MR, Adams WK, Kinzel GL. A finite element survey of eleven endosseous implants[J]. J Prosthet Dent, 1990, 63 (4): 457-465. 被引量:1
  • 9兰泽栋,林珠,龙红月,周磊.支抗种植体长度对骨界面应力分布的影响[J].中国口腔种植学杂志,2003,8(3):112-115. 被引量:22
  • 10朱胜吉,荣起国,周彦恒.微螺钉型种植体支抗长度及直径对应力分布影响的三维有限元研究[J].口腔正畸学,2006,13(2):49-52. 被引量:25

二级参考文献49

共引文献83

同被引文献44

  • 1赵弘,刘洪臣,顾晓明.骨皮质厚度及植入角度对支抗种植钉影响的三维有限元分析[J].口腔颌面修复学杂志,2010,11(1):24-27. 被引量:14
  • 2张扬,张丹,冯翠娟.微小种植体正畸支抗生物力学的三维有限元分析[J].上海口腔医学,2005,14(3):281-283. 被引量:33
  • 3邓锋,张磊,张翼,宋锦璘,樊瑜波.微植体支抗-骨界面的生物力学研究及微植体颈部优化设计探讨[J].四川大学学报(医学版),2007,38(4):701-704. 被引量:10
  • 4Kuroda S,Yamada K, Deguchi T, et al. Root proximity is a majorfactor for screw failure in orthodontic anchorage [J]. Am J OrthodDentofacial Orthop, 2007,131 :S68-S73. 被引量:1
  • 5Singh S, Mogra S,Shetty VS, et al. Three -dimensional finiteelement analysis of strength, stability, and stress distribution inorthodontic anchorage: a conical, self -drilling miniscrew implantsystem[J]. Am J Orthod Dentofacial Orthop,2012,14⑶:327-336. 被引量:1
  • 6Jang HJ, Kwon SY, Kim SH, et al. Effects of washer on the stressdistribution of mini -implant: A finite element analysis [J].AngleOrthod, 2012,82(1): 137-144. 被引量:1
  • 7Liu TC,Chang CH, Wong TY, et al. Finite element analysis ofminiscrew implant used for orthodontic anchorage [J].Am J OrthodDentofacial 0rthop,2012,141(4):468-476. 被引量:1
  • 8Favero L, Giagnorio C,Cocilovo F. Comparative analysis ofanchorage systems for micro implant orthodontics [J]. Prog Orthod.2010,11(2):105-117. 被引量:1
  • 9Barros SE, Janson G, Chigueto K, et al. Effect of mini -implantdiameter on fracture risk and self-drilling efficacy [J].Am J OrthodDentofacial Orthop, 2011,140(4):el81-el92. 被引量:1
  • 10M isch CE Degidi M.Five-Year Prospective Study of Immediate/ Early Loading of Fixed Prostheses in Completely Edentulous Jaws with a Bone Quality-Based Implant System[J].Clinical Implant Dentistry and Related Research,2003 (5):17-27. 被引量:1

引证文献6

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部