摘要
采用流动注射分析仪测定渔业养殖海水中的挥发酚。方法的检出限为0.89μg/L;线性范围为0.002~0.1 mg/L,线性相关系数r≥0.999;其相对标准偏差为1%~5%,样品测定频率为18样/h。本方法操作简便、灵敏度较高,线性范围、精密度和准确度均符合国家标准。对海水水样加标回收率的测定为90%~110%,结果令人满意。
The volatile phenolics in seawater were determined by the flow injection analysis-4-AAP method with automated flow injection analyzer developed by Titan Instruments.The detection limit was 0.89 μg/L and liner range was 0.002~0.1 mg/L(r≥0.999).The relative standard deviation was 1%~5% and recovery for seawater was 90%~110%.The method is rapid,simple and sensitive.The linear range,precision and accuracy all consist with GB standard.
出处
《海洋环境科学》
CAS
CSCD
北大核心
2011年第1期102-104,共3页
Marine Environmental Science
基金
国家科技部"十五"科技攻关项目(2004BA210A05)
关键词
挥发酚
流动注射
在线蒸馏
flow injection analysis; in-line distillation; volatile phenolics