冠脉支架膨胀过程的有限元模拟 被引量:4

Finite Element Simulation of Coronary Stent Expansion Process
下载PDF
导出
摘要 在研究确定了冠脉支架的几何模型、有限元模型、材料模型、边界条件、求解控制的基础上,利用有限元法对冠脉支架的膨胀过程进行了模拟分析,并对分析结果进行了实验验证。结果表明,利用有限元法建立的力学模型可以在不考虑球囊和血管作用的情况下,较好地模拟和分析冠脉支架的实际膨胀过程。通过该方法设计制造的冠脉支架长度缩短率和狗骨头率分别达到1.31%和3.66%,与有限元分析结果吻合良好。该方法可为新型结构冠脉支架的优化设计提供依据。 Based on the research of geometry model, finite element model, material model, boundary conditions and solution control of coronary stent, coronary stent expansion process simulation analysis is made using finite element method, and experimentally validated to the analysis results. It shows that the mechanics model by finite element method can simulate and analyze the real expansion process of coronary stent well under the condition of ignoring the action between saeeulus and blood vessel. The coronary stent expansion rate and dog bone rate made by this method reaches 1.31% and 3.66% respectively, which meets the analysis results very well. So this method provides technical basis for new construction coronary stent.
出处 《材料导报》 CSCD 北大核心 2009年第20期70-72,97,共4页 Materials Reports
基金 吉林省科技厅资助基础研究项目(20040509)
关键词 冠脉支架 有限元模拟 膨胀 coronary stent, finite element simulation, expansion
  • 相关文献

参考文献11

二级参考文献58

共引文献50

同被引文献37

  • 1顾兴中,倪中华.微孔结构血管支架的激光切割工艺[J].华中科技大学学报(自然科学版),2007,35(S1):143-146. 被引量:10
  • 2张艺浩,李红霞,王希诚.血管支架疲劳寿命分析[J].哈尔滨工业大学学报,2011,43(S1):86-90. 被引量:11
  • 3李红霞,王希诚.基于不同扩张模拟方式的支架优化设计[J].哈尔滨工业大学学报,2011,43(S1):267-272. 被引量:4
  • 4Belforte D A. Milestones in industrial laser materials pro- cessing[J] Ind I.aser Solut, 2010,250 (10) : ] 6. 被引量:1
  • 5Muhammad N, Whitehead D, Poor A, et al Comparison of dry and wet fibre laser profile cutting of thin 316I. stainless steel tubes for medical device applications [J]. J Mater Process Techn, 2010,210(15) : 2261. 被引量:1
  • 6Meng H Y, Liao J H, Zhou Y H, et al. I.aser micro-pro- cessing of cardiovascular stent with fiber laser cutting sys- tem[J]. Opt I.aser Techn,2009,41(3):300. 被引量:1
  • 7谢冀江,祁磊.冠脉支架的长度缩短率和狗骨头率分析[J].光学精密工程,2008,16(1):87-91. 被引量:5
  • 8Rogers C, Tseng DY, Squire JC, et al. Balloon-artery interactions during stent placement: a finite element analysis approach to pressure, compliance, and stent design as contributors to vascular injury [ J]. Circulation Research, 1999, 84(4) :378 -383. 被引量:1
  • 9Kiousis DE, Wulff AR, Holzapfel GA. Experimental studies and numerical analysis of the inflation and interaction of vascularballoon catheter-stent systems [ J ]. Annals of Biomedical Engineering, 2009, 37 (2) :315 - 330. 被引量:1
  • 10Lim D, Cho SK, Park WP, et al. Suggestion of potential stent design parameters to reduce restenosis risk driven by foreshortening or dogboning due to non-uniform balloon-stent expansion [ J]. Annals of Biomedical Engineering, 2008, 36 (7) :1118 - 1129. 被引量:1