显微CT骨标本扫描的伪影评估与消减控制 被引量:9

Evaluation and reducing control of the artifacts of bone specimens scanned by Micro-CT
下载PDF
导出
摘要 目的观察分析显微CT骨标本扫描的伪影强度及对周边正常骨组织的影响程度,寻求消减、控制伪影及骨组织有效测试方法。方法选取皮质骨、牙体及不同种类、不同弹性模量的内置钉骨标本为扫描对象,以不同放置条件、不同扫描协议、不同校正选项扫描,观察、测试重建后的伪影程度。结果高密度物质产生的伪影对骨组织形态、密度分析影响较大。选择适当标本放置、360度加强型扫描协议及伪影消减对伪影均具有一定的消减与控制作用,采用校样同扫校正,能够达到周边骨组织的有效分析测试。结论显微CT对含高密度物质标本的扫描分析,应尽可能地消减、控制伪影,重新校正周边骨组织密度后,才能达到骨组织的有效分析测试。 Objective To observe and analyze the bony artifacts in micro-CT scanning and their impact on normal bone tissue, and to find an effective method to reduce and control the bony artifacts and properly measurement of bone tissues. Methods Cortical bone, tooth and internal nails of different kinds and elastic modulus were scanned under diverse setting conditions, scanning protocols and modifying selections. The artifacts were observed and evaluated after reconstruction. Results The influences of artifacts of specimen with high density were much higher on the exhibition of bony appearance and density analysis. The artifacts were decreased and controlled by the improvement of setting conditions and 360° strengthen scanning protocol. Effective test could be realized by co-scanning correction. Conclusion The artifact should be reduced as much as possible when a specimen with high density is scanned by micro-CT. In order to obtain effective analysis results, bony tissue density should be re-corrected.
出处 《中国体视学与图像分析》 2009年第1期85-92,共8页 Chinese Journal of Stereology and Image Analysis
基金 国家863课题(2007AA02Z468)
关键词 MICRO-CT 伪影 骨形态计量 micro-CT artifacts bone morphometry
  • 相关文献

参考文献7

二级参考文献29

  • 1张慧滔,张朋,宋黎明.硬X射线调制望远镜直接解调成像方法快速实现[J].CT理论与应用研究(中英文),2005,14(1):17-23. 被引量:1
  • 2卢方军,李惕碚,吴枚,孙学军.ME/EXOSAT银道扫描数据的直接解调成像[J].天文学报,1997,38(1):56-66. 被引量:6
  • 3Partial volume artifact reduction in computed tomography[C]. Jiang Hsieh, Applied Science Laboratory, G.E. Medical Systems, Proc. IEEE Intern. Conf. On image Processing,1996, 567-570. 被引量:1
  • 4Nonlinear Partial volume artifact correction in Helical CT[C]. Jiang Hsieh, Applied Science Laboratory, G.E. Medical Systems, Proc. IEEE Intern. Conf. On image Processing, 1999, 2035-2039. 被引量:1
  • 5Cause and correction for 3D Image Artifact in Helical CT[J]. Jiang Hsieh, Applied Science Laboratory, G.E. Medical Systems, Proc. IEEE Intern. Conf. On image Processing, 1999,2040-2044. 被引量:1
  • 6Reduction of mental streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm[C]. B.De Man etc., IEEE Intern. Conf. On image Processing, 2000,850-854. 被引量:1
  • 7Ge Wang etc.,Iterative Deblurring for CT Metal artifact reduction[J]. IEEE Transactions on medical imaging, 1996,Vol.15, NO.5, October, 657-664. 被引量:1
  • 8Chye Hwang Yan etc., Reconstruction Algorithm for polychromatic CT Imaging: Application to Beam Hardening Correction[J]. IEEE Transactions on medical imaging, 2000,Vol.19, NO.1. 被引量:1
  • 9Thomas H. Newton, M.D., D. Gordon Potts, M.D.Radiology of the skull and brain technical aspects of computed tomography[J ]. 1981, 3956-3992. 被引量:1
  • 10Gabor T Herman.Image reconstruction from Projections[M].Academic Press,INC,1980.64-66. 被引量:1

共引文献59

同被引文献95

引证文献9

二级引证文献38

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部