摘要 研究了基于Matrox图像采集卡的DR图像采集中的信号控制问题;设计了平板状态控制和数据处理流程;提出了原始图像的隔行扫描及其区域交换算法;解决了如何将原始图像调整为规整的X线图像的问题,并在此基础上设计了DR图像的采集、显示和存储程序,分别对水模和人体部位进行了X线摄影实验。结果表明,本文提出的方法准确可靠,解决了DR图像采集过程中的信号控制和图像处理的问题,获取的X线数字图像真实有效,已经被应用于实际的DR系统。 The signal control strategy in a Matrox Meteor Ⅱ /Camera link card-based DR (Direct digital radiography) image acquisition system is studied. The processing flow for status control and data process of the flat panel is studied. A line interval scan and a area exchange algorithm are presented to process the original image data acquired from the detector array. The problem on the adjustment of the abnormal original image is solved and a program is developed for DR image acquisition, display and storage. Experimental results on water models and human bodies show that the method are correct and digital X ray images are valid. The method is applied in the clinical practices of the direct digital radiography system.
出处 《数据采集与处理》 CSCD 北大核心 2008年第3期250-253,共4页 Journal of Data Acquisition and Processing
关键词 直接数字化X线摄影 图像采集 信号控制 医学图像处理 direct digital radiography image acquisition signal control medical image processing
  • 相关文献

参考文献6

  • 1郗雁平.数字化X线成像技术[J].CT理论与应用研究(中英文),2004,13(2):30-32. 被引量:3
  • 2Aach T, Schiebel U, Spekowius G. Digital image acquisition and processing in medical X-ray imaging [J]. Journal of Electronic Imaging, 1999,8(1) : 7-22. 被引量:1
  • 3Kasap S O, Rowlands J A. Direct-conversion flatpanel X-ray image detectors [J]. lEE Proceedings: Circuits, Devices and Systems, 2002,149 (2) : 85-96. 被引量:1
  • 4唐彬,夏明.基于Matrox Pulsar图象采集卡研发的图象处理系统[J].微计算机信息,2004,20(4):41-42. 被引量:8
  • 5王宾 潘建寿 梁雁冰.基于Matrox卡的视频图像运动检测[J].西北大学学报,2004,2(3):1-7. 被引量:6
  • 6Zhao Wei, Blevis Ira, Germman S, et al. A flat panel detector for digital radiology using active matrix readout of amorphous selenium [J]. Proceedings of SPIE-The International Society for Optical Engineering, 1996, 2708: 523-531. 被引量:1

二级参考文献2

  • 1庄益明.CCD医用X线影像增强电视的发展[J].世界医疗器械,1997,3(3):11-15. 被引量:1
  • 2Matrox Electronic Systems Ltd. Matrox Pulsar Installation and Hardware Reference. CANADA. 1998. 被引量:1