期刊文献+

用于探测生物芯片的制冷型ICCD系统

Cooled ICCD system for biochip detection
下载PDF
导出
摘要 利用光锥耦合的ICCD系统探测荧光染料标记的生物芯片,并对CCD芯片和像增强器制冷,以提高探测灵敏度。基于实验分析结果,指出背景噪声的主要来源为杂散光和生物芯片基底所发的荧光,指出用镜头成像限制了系统探测灵敏度的提高,可采用低荧光物质作为生物芯片的基底对系统加以改进。 An optical cone-coupled ICCD system was developed to detect the biochips labeled with fluorescent molecules. The CCD chip and image intensifier were cooled to increase the sensitivity of the system. Based on the analyzed results of the experiment, it is determined that the major background noise comes from the stray light and the fluorescence from the substrate impurities of the biochip, and the sensitivity of the system is limited by the objective lens for imaging. The background noise can be effectively reduced and the detection sensitivity can be improved by using a substrate with less fluorescent impurities, such as a quartz substrate.
作者 王红球 蒋硕
机构地区 清华大学物理系
出处 《应用光学》 CAS CSCD 2008年第3期339-342,共4页 Journal of Applied Optics
基金 清华大学基础研究基金(052205006)
关键词 光学仪器 生物芯片 ICCD 背景噪声 optical instrument biochip ICCD background noise
  • 相关文献

参考文献9

二级参考文献13

  • 1[2]LAPRADE B N,REINHART S T,WHEELER.A low noise figure microchannel plate optimized for Gen Ⅲ image intensification systems[J].SPIE,1990,1243:162-172. 被引量:1
  • 2[3]FELLER W B,COOK L M,FRASER G W,et al.Low noise microchannel plates[J].SPIE,1989,1072:138-148. 被引量:1
  • 3[4]ZHONG W J S,WILLIAMS J.Etchable core glass compositions and method for manufacturing a high performance microchannel plate:USA,5108961[P].1992-04-28. 被引量:1
  • 4MkGVN992-1996, Image intensifier tubes speci-fications philips Photonics[S].. 被引量:1
  • 5S P A Fodor, R P Rava, X C Huang, A C Pease, C P Holmes, C L Adams. Multiplexed biochemical assays with bilolgical chips[J], Nature, 1993,364: 555~556. 被引量:1
  • 6M Schena, D Shalon, R W Davis, P O Brown. Quantitative monitoring of gene expression patterns with a complementary DNA microarry [J], Science, 1995, 270: 467~470. 被引量:1
  • 7J Montagu, P Honkanen. Fluorescence ArrayScanner[J], SPIE, 1999, 3779, 284~292. 被引量:1
  • 8Bruce N Laprade,et al. A low noise figure micro-channel plate optimized for Gen Ⅲ image intensification systems[J]. SPIE,1990,1243:162-172. 被引量:1
  • 9Ferguson J A , Steermer F J , Walt D R . High-density fiber-optical DNA random microsphere array.Anal Chem,2000 ,72(27):5618-5624 被引量:1
  • 10Mathies R A,Peck K. Optimization of high-sensitivity fluorescence detection.Anal Chem,1990,62(17):1786-1791 被引量:1

共引文献95