摘要
对市售的膨化食品、鲜、干水产品、豆制品采用微波仪消解,在电感耦合等离子体质谱仪上采用钪(Sc)作为内标元素对其中铝含量进行直接定量分析。方法前处理简便,测量快速、灵敏、准确。方法检出限为0.3mg/kg,相对标准偏差〈3%,加标回收率在85%-104%。
The determination of Aluminium in foods with microwave digestion - inductivly coupled plasma mass spectrometer ( ICP - MS) were introduced in this paper, using the scandium (Se) as internal standard. This method was simple , rapid, sensitive and accurate. The determination limit was 0.3 mg/kg,the recovery rate of standand addition was 85 % - 104 % and the relative standard deviation was less than 3 %.
出处
《辽宁化工》
CAS
2008年第1期68-69,72,共3页
Liaoning Chemical Industry