摘要 本文对扫描电声显微镜电声成像中出现的干扰信号以及假信号的产生进行了讨论.指出扫描电声显微镜设备周围存在的高强度电场、磁场、高频设备和仪器本身的接地是造成干扰信号出现的主要原因,同时给出了一种能有效防止以上干扰的样品一探测器组件结构.而在电声像中假信号产生的原因则主要是由于样品和探测器接地不当所引起.在进行电声成像观察和记录时最好关掉电镜的对比度旋钮. The effects of disturbed and false signals on Scanning Electron Acoustic Microscope (SEAM)are presented in this paper. Various reasons such as high electric fields, magnetic fields and high frequency instruments causing the disturbed signals on SEAM are discussed and a sample of detection assembly is give out, simultaneously. Emergence of false signals in SEAM is also analyzed on the bases of the upper surface of the sample earthed and the sample-transducer interface unearthed, the upper surface of the transducer earthed and the surface of the sample unearthed, and with both the upper surface of the sample and the transducer earthed. It would be best that the contrast switch is turned off when electron acoustic images are observed and imaged.
出处 《无机材料学报》 SCIE EI CAS CSCD 北大核心 1997年第4期637-640,共4页 Journal of Inorganic Materials
基金 "63"高技术!(715-10-02-03) 国家自然科学基金!59372115
关键词 干扰信号 假信号 扫描电声显微镜 电子显微镜 disturbed signal, false signal, scanning electron acoustic microscope(SEAM)
  • 相关文献

参考文献5

  • 1殷庆瑞,唐振,张惠静,江福民,王通,王惠芳,丁积祥,梁坚,韦建勇,朱祖福.扫描电—声显微镜及其应用[J].电子显微学报,1990,9(4):53-61. 被引量:6
  • 2殷庆瑞,Prceedings of First Asian Meeting on Ferroelectrics,1995年 被引量:1
  • 3张冰阳,Proceedings of First Asian Metting on Ferroelectrics,1995年 被引量:1
  • 4张冰阳,Proceedings of First Southeast Asian Symposium on Thermophysical Properties,1995年 被引量:1
  • 5殷庆瑞,Ferroelectrics,1994年,151卷,97页 被引量:1

二级参考文献3

  • 1梁坚,电子显微学报,1988年,7卷,2期,66页 被引量:1
  • 2殷庆瑞,自然杂志,1985年,8卷,12期,345页 被引量:1
  • 3殷庆瑞,物理,1984年,4期,246页 被引量:1

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部