应用改良的FS-3D-TrueFISP序列诊断膝关节软骨缺损性病变 被引量:5

Application of improved FS-3D-TrueFISP sequence in diagnosis of hyaline cartilage defects in knee joint
下载PDF
导出
摘要 目的以关节镜为金标准,评价经过改良的脂肪抑制三维真实稳态自由进动序列(FS-3D-TrueFISP)对于膝关节软骨缺损性病变的诊断价值。方法采用经过改良的FS-3D-TrueFISP序列及脂肪抑制三维快速小角度激发序列(FS-3D-FLASH)对21例膝关节病变患者进行检查,其中13例患者同时行脂肪抑制快速自旋回波质子密度加权序列(FS-FSE-PDWI)扫描,将各序列诊断结果与关节镜分级诊断结果相对照。结果与关节镜结果相对照FS-3D-TrueFISP序列对于软骨缺损行病变的敏感性、特异度及表示准确度的Kappa值分别为83.3%、99.2%、0.77;FS-3D-FLASH序列相应各值分别为80.5%、97.7%、0.68;FS-FSE-PDWI序列分别为57.1%、97.6%、0.60。结论经过合理改良后的FS-3D-TrueFISP序列对于膝关节软骨缺损性病变具有较高的敏感性,应用这一技术有助于膝关节软骨缺损性病变的诊断。 Objective To evaluate the value of improved fat-suppressed-3D-true fast imaging with steady-state precession (FS-3D-TrueFISP) sequence for detecting hyaline cartilage defects of knee joint, using arthroscopy as the gold standard. Methods Twenty-one patients were imaged with fat-suppressed-3D-fast low angle shot (FS-3D-FLASH) sequence and FS- 3D-TrueFISP sequence with improved imaging parameters, and 13 patients imaging with fat-suppressed-fast spin echo-proton density weight image (FS-FSE-PDWI) sequence at same time, then compared the imaging findings with arthroscopy. Results Comparing with arthroscopy, the sensitivity, specificity and Kappa were 83.3%, 99.2% and 0. 77 respectively with FS- 3D-TrueFISP; 80. 5%, 97.7% and 0.68 respectively with FS-3D-FLASH sequence; 57.1%, 97.6% and 0. 60 respectively with FS-FSE-PDWI sequence. Conclusion The improved FS-3D-TrueFISP sequence has higher sensitivity for detecting hyaline cartilage defects in knee joint. Use of this technique may strengthen the role of MR imaging for detecting hyaline cartilage defects.
出处 《中国医学影像技术》 CSCD 北大核心 2007年第1期105-108,共4页 Chinese Journal of Medical Imaging Technology
基金 新世纪优秀人才支持计划(NCET)资助(985-2-070-113)。
关键词 软骨 关节 磁共振成像 膝损伤 Cartilage, articular Magnetic resonance imaging Knee injuries
  • 相关文献

参考文献11

  • 1Cooper C.Occupational activity and the risk of osteoarthritis[J].J Rheumatol,1995,43(1):10-12. 被引量:1
  • 2李松柏,何翠菊,吴振华,白希壮,孙文阁,李春奎,戚喜勋,李延亮,徐克.膝关节软骨缺损性病变的MR成像序列研究[J].中华放射学杂志,2003,37(11):967-972. 被引量:34
  • 3孙燕萍,杜湘珂,周殿阁,张斌.三维脂肪抑制快速毁损梯度回波序列对髌股关节软骨病变的评价[J].中国医学影像技术,2002,18(9):930-932. 被引量:5
  • 4Recht MP,Piraino DW,Paletta GA,et al.Accuracy of fat-suppressed three-dimensional spoiled gradient-echo FLASH MR imaging in the detection of patellofemoral articular cartilage abnormalities[J].Radiology,1996,198(1):209-212. 被引量:1
  • 5Scheffler K.A pictorial description of steady states in fast magnetic resonance imaging:concept[J].Magn Res Med,1999,11 (2):291-304. 被引量:1
  • 6Scheffler K.Henning J.Is TrueFISP a gradient-echo or a spineecho sequence[J] ? Magn Res Med,2003,49(2):395-397. 被引量:1
  • 7Huang TY,Huang IJ,Chen CY,et al.Are TrueFISP images T2/T1-weighted[J]:Magn Res Med,2002,48(4):684-688. 被引量:1
  • 8Weckbach S,Mendlik T,Horger W,et al.Quantitative assessment of patellar cartilage volume and thickness at 3.0 Tesla comparing a 3D-fast low angle shot versus a 3D-true fast imaging with steady-state precession sequence for reproducibility[J].Invest Radiol,2006,41(2):189-197. 被引量:1
  • 9Disler DG,McCauley TR,Kelman CG,et al.Fat-suppressed three-dimensional spoiled gradient-echo MR imaging of hyaline cartilage defects in the knee:comparison with standard MR imaging and arthroscopy[J].AJR,1996,167(1):127-132. 被引量:1
  • 10Brown TR,Quinn SF.Evaluation of chondromalacia of the patellofemoral compartment with axial magnetic resonance imaging[J].Skeletal Radiol,1993,22(5):325-328. 被引量:1

二级参考文献16

  • 1[1]Suh JS,Cho JH,Shin KH,et al. Chondromalacia of the knee:evaluation with a fat-suppression three-dimensional SPGR imaging after intraenous contrast injection[J].JMRI,1996,6(6):884-888. 被引量:1
  • 2[2]Recht MP,Kramer J,Marcelis S,et al. Abnormalities of articular cartilage in the knee:analysis of available MR techniques[J].Radiology,1993,187(2):473-478. 被引量:1
  • 3[3]Sittek H,Eckstein F,Gavazzeni A,et al. Assessment of normal patellar cartilage volume and thickness using MRI:an analysis of currently available pulse sequences[J].Skeletal Radiol,1996,25(1):55-62. 被引量:1
  • 4[4]Recht MP,Piraion DW,Paletta GA,et al. Accuracy of fat-suppressed three-dimensional spoiled gradient-echo FLASH MR imaging in the detection of patellofemoral articular cartilage abnormalities[J].Radiology,1996,198(1):209-212. 被引量:1
  • 5[5]Disler DG,Peters TL,Muscoreil SJ,et al. Fat-suppressed spoiled GRASS imaging of knee hyaline cartilage:technique optimization and comparison with conventional MR imaging[J].AJR,1994,163:887-892. 被引量:1
  • 6[6]Disler DG. Fat-suppression three dimensional spoiled gradient-recalled MR imaging:assessment of articular and physeal hyaline cartilage[J].AJR,1997,169(4):1117-1123. 被引量:1
  • 7[7]Mankin JM,Thrasher AZ. Water content and binding in normal and osteoarthritic Human cartilage[J].J Bone Jiont Surg,1975,57(1):76-80. 被引量:1
  • 8[8]Hodler J,Berthiaume MJ,Schweitzer ME,et al. Knee joint hyaline cartilage defects:a comparative study of MR and anatomic sections[J].J Comput Assist Tomogr,1992,16(4):597-603. 被引量:1
  • 9[9]Rubenstein JD,Kim JK,Morava-Protner I,et al.Effects of collagen orientation on MR imaging characteristics of bovine articular cartilage[J].Radiology,1993,188(1):219-226. 被引量:1
  • 10[10]Hayes CW,Sawyer RW,Conway WF. Patellar cartilage lesions:in vitro detection and pathologic correlation[J].Radiology,1990,176(2):479-483. 被引量:1

共引文献37

同被引文献41

引证文献5

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部