期刊文献+

尿素氮-葡萄糖双功能分析仪的研究 被引量:3

Study on Bifunctional Analyser of Urea-nitrogen and Glucose
下载PDF
导出
摘要 由固定化脲酶、谷氨酸脱氢酶、谷氨酸氧化酶、葡萄糖氧化酶的复合酶膜组成的双电极系统,可以同时测定尿素氮和葡萄糖的含量,每次进样量为25μl,20s即可测定出尿素氮和葡萄糖的含量。在0~60mg/dl尿素氮、0~500mg/dl葡萄糖范围内具有良好的线性关系。连续测定20次的变异系数分别为1.02%和1.05%。酶膜使用寿命为两星期以上。此仪器可广泛应用于临床检验和体育训练中。 The urea-nitrogen and glucose bifunctional analyser consisted of two electrodes, immobilized urease, glutamante oxidase, glutamate dehydrogenase and glucose oxidase can simultaneously determine the concentration of urea-nitrogen and glucose. The linear range is 0-60mg/dl for urea-nitrogen and 0 - 500mg/dl for glucose with a response of 20 seconds and 25ul sample injection. The coefficient of variation is 1. 02% and 1. 05% respectively in 20 assays. The enzyme membrances can be used continuously at 25C for two weeks. This bifunctional analyser can be used widely to determine urea-nitrogen and glucose at the same time in clinical and physical training.
出处 《生物工程学报》 CAS CSCD 北大核心 1996年第2期189-193,共5页 Chinese Journal of Biotechnology
关键词 尿素氮 葡萄糖 双功能分析仪 生化检验 Enzyme electrode, urea-nitrogen, glucose
  • 相关文献

参考文献8

  • 1冯德荣,吉林大学自然科学学报,1993年,增刊,80页 被引量:1
  • 2冯德荣,食品与发酵工业,1993年,4卷,33页 被引量:1
  • 3钟丽婵,生物工程学报,1992年,8卷,87页 被引量:1
  • 4朱忠勇,实用医学检验学,1992年 被引量:1
  • 5冯德荣,山东科学,1990年,3卷,1页 被引量:1
  • 6冯德荣,中国医药工业杂志,1990年,21页 被引量:1
  • 7黎高翔,全国固定化生物催化剂学术讨论会论文摘要,1988年 被引量:1
  • 8冯德荣,全国第二次工业生化学术会议论文摘要,1987年 被引量:1