摘要
经试验,在温度45~56℃、相对湿度60%~85%的条件下,以750mg/L环氧乙烷作用5小时,可将宫内节育器灭菌。用无菌空气气洗2~5次,可将节育器上残留环氧乙烷量减至5ppm以下。
It was found in experiment that 750 mg/L ethylene oxide with a 5-hr contact time at 45~56℃ and RH 60% could sterilize the intrauterine contraceptive devices and washing with sterile air for 2-5 times could reduce the residual ethylene oxide on the contraceptive device to lower than 5 ppm.
出处
《中国消毒学杂志》
CAS
北大核心
1996年第2期101-103,共3页
Chinese Journal of Disinfection
关键词
避孕器
灭菌
宫内节育器
环氧乙烷
sterilization of contraceptive device intrauterine cotraceptive device ethylene oxide