摘要
作者对Gamma钉系统做了如下主要的改进:(1)钉的尺寸改小以适合国人,原髓内棒上段弯度取消,改为纵内切斜面8°,颈螺钉改为真正的拉力钉,远瑞交锁螺钉改用1枚;(2)瞄准器由多部件接合体改为半固定结合体,设有校准程序;(3)采用改良三点显露法使开放复位、安装髓内棒和颈螺钉在同一个小切口内完成;(4)术中不用放射性检查或仅在确定颈螺钉深度时使用。采用改良Gamma钉手术治疗73例股骨转子周围部骨折,术后随诊5~38个月,通过放射学及临床评价,疗效满意。此法有提高复位内固定质量,减少并发症,允许早期负重,不用顾虑术者接受累积放射性损害的优点。
AbstractThe authors has modified the Gamma nailing sys-tem for the treatment of peritrochanteric fracture of thefemur mainly on the following points:(1)the size ofthe nail is varied to be adjustable to the femur of theChinese patients;the curvature of intramedullary rod iseliminated, the cervical screw is changed into a true lagscrew and instead of two, only one distal lockingscrews is used.(2)the targeting device is changed to anewly designed semi-fixed combination.(3)a smallerincision is used for open reduction and the insertion ofthe intramedullarv rod and the cervical screw.(4)x-rayscreening is used only for intraoperative confirmation ofthe depth of the cervical screw.73 cases of per-itrochanteric fracture of the femur were so treated andthe patients were followed up for 5~38 months. Theclinical and radiographical results were all satisfactory.
出处
《中华骨科杂志》
CAS
CSCD
北大核心
1995年第8期512-515,共4页
Chinese Journal of Orthopaedics