期刊文献+

电感耦合等离子体发射光谱法(ICP-AES)测定污水中总磷 被引量:8

Determination of Total Phosphorus in Sewage by ICP-AES
下载PDF
导出
摘要 采用美国热电公司TraceScan型端视ICP AES光谱仪,对测定污水中总磷的最佳测试条件的选择进行了探讨,并用所选取的最佳测试条件进行测试。结果表明,本方法简便,可以直接上机进行测定;快速而准确,相对误差为0%~2.00%;精密度好,相对标准偏差为0.3%~4.2%(n=3);回收率在104.00%~118.00%之间,检出限0.015mg/L。 The optimal condition of testing total phosphorus amount of sewage was studied and the tests were carried out according to the selected optimal condition by means of Trace Scan type ICP-AES spectrograph made by ThermoElemental Company in the USA. The test results showed that this method of testing total phosphorus amount of sewage by means of ICP-AES technology is not only simple and convenient but also fast and accurate. Its relative error is between 0%~2.00%, and relative standard deviation is between 0.3%~4.2%(n=3). Its recovery rate is between 104.00%~118.00%. Its detection limit is 0.015mg/L.
出处 《上海农业学报》 CSCD 2005年第2期16-18,共3页 Acta Agriculturae Shanghai
  • 相关文献

参考文献3

  • 1戴特AR 李金英等(译).电感耦合等离子体质谱分析的应用[M].北京:原子能出版社,1998.. 被引量:5
  • 2叶伊兵.中国环境保护标准汇编[S].北京: 中国标准出版社,2000.188-191. 被引量:1
  • 3陈新坤.电感耦合等离子光谱原理与应用[M].天津:南开大学出版社,1987.22-50. 被引量:1