胫骨定量超声测量及临床应用探讨 被引量:2

Quantitative ultrasound measurement of tibia and clinical application
下载PDF
导出
摘要 用SoundScan2000骨定量超声(QUS)仪测量胫骨超声速度(SOS),同时与单光子吸收法(SPA)测量前臂1/3处骨矿密度(BMD)比较。两方法测得208例患者和健康志愿者结果相关(r=0.678,P<0.001)。42例健康绝经妇女SOS和BMD与绝经时间呈负相关(r=-0.417和-0.479,P<0.01),73例>40岁的健康志愿者SOS和BMD与年龄呈负相关(r=-0.293和-0.373,P<0.05)。与性别和年龄相匹配的正常参考值比较,结果<x-2s者QUS检出34例,占16.3%;SPA23例,占11.0%,QUS的诊断敏感度约是SPA的1.5倍。 We have assessed a new method of quantitative ultrasound (QUS,with SoundScan 2000 of Myriad Ultrasound Systems Ltd)for measuring the speed of sound(SOS,m s-1) and diagnosis of osteoporosis in the conical bone of the right tibial shaft. The QUS was compared with single photon absorptiometry(SPA) for measuring the bone mineral density (BMD, g cm-2) at the forearm (1/3 site).SOS significantly correlated with BMD (r= 0. 678, P<0. 001 ) in 208 subjects, including patients with various metabolic bone diseases and 99 healthy volunteers by QUS and SPA. The QUS was most reproducible with intra- and intercoefficients of 0. 22% and 0. 39% respectively. In the healthy late postmenopausal women,there was a correlation of SOS and BMD with years since menopause (r =-0. 417 and -0. 479,P<0. 01 and 0. 002). In the healthy volunteers aged over 40 years,there was also a correlation of SOS and BMD with age(r=-0. 293 and -0. 373,P<0. 05 and 0. 01 ). Compared with healthy subjects matched by sex and age,34 cases by QUS and 23 cases by SPA were were detected to have the values lower than x-2S of healthy subjects,the diagnostic sensitivity of QUS to SPA being about 1. 5 times. The results show that the technique of QUS is highly sensitive and precise. The diagnosis of osteoporosis and prediction of osteoporotic fracture by QUS are satisfactory.
出处 《中国骨质疏松杂志》 CAS CSCD 1997年第1期43-45,共3页 Chinese Journal of Osteoporosis
关键词 胫骨 定量超声 超声测量 骨质疏松 单光子吸收法 Quantitative ultrasound Single photon absorptiometry Osteoporosis
  • 相关文献

同被引文献8

引证文献2

二级引证文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部