期刊文献+

四电极电压补偿测量质子交换膜的导电率 被引量:1

The Conductivity Measure of Proton-Exchange Membrane with Four Electrode Voltage Compensation
下载PDF
导出
摘要 根据质子交换膜导电的特点,对其电导率的测试方法进行了研究,分析了通常采用的直流电、交流阻抗以及同轴探针测试方法在测量膜电导率时容易产生误差的原因。并设计了一种四电极电压补偿测试系统,能有效地避开电极极化以及仪器高频感应的影响,使测量结果精确可靠。 The various principles of intermediaries to measure the conductivity of proton exchange membrane such as direct current measuring, ac impedance spectroscopy and open-ended coaxial probe method according to the characteristics of proton exchange membrane were shown and studied respectively. The reason of error resulting from the measuring methods was analyzed .A circuit testing system with four electrode voltage compensation was designed and described, which can effectively eliminate the influence of significant electrode polarization at low frequencies due to interfacial impedance and component of instrumental response at high frequencies. The measuring results of the conductivity of proton exchange membrane are not only precise but also reliable.
出处 《实验室研究与探索》 CAS 2004年第8期22-25,共4页 Research and Exploration In Laboratory
基金 东风汽车公司燃料电池电动汽车项目资助。
关键词 质子交换膜 电导率 电压补偿 误差 proton exchange membrane conductivity voltage compensation error
  • 相关文献

参考文献12

  • 1林维明主编..燃料电池系统[M].北京:化学工业出版社,1996:205.
  • 2衣宝廉.燃料电池[M].北京:化学工业出版社,2000.. 被引量:21
  • 3Chuy C, Basura V I., Simon E, et al. Electrochemical characterization of ethylenetetrafluoroethylene-g-polystyrenesulfonic acid solid polymer electrolytes[J].Journal of The Electrochemical Society, 2000,147(12):4453-4455. 被引量:1
  • 4Mohr R, Kudela V, Schauer J, et al. Comparison of different cells for resistance determination of freely standing polymer membranes developed for direct methanol fuel cell(DMFC) applications[J].Desalination, 2002,(147):191-196. 被引量:1
  • 5Gardner C.L, Anantaraman A.V. Measurement of membrane conductivities using an open-ended coaxial probe[J].Journal of Electroanalytical Chemistry, 1995,(395):67-73. 被引量:1
  • 6Beattie P D, Orfino F P, Basura V I, et al. Ionic conductivity of proton exchange membranes[J].Journal of Electroanalytical Chemistry, 2001,(503):45-56. 被引量:1
  • 7Orfino F P, Holdcroft S. The morphology of Nafion: are ion cluster bridged by channels or single ionic sites?[J].Journal of New Materials for Electrochemical Systems, 2000,(3):285-290. 被引量:1
  • 8Costamagna P, Srinivasan S. Quantum jumps in the PEMFC sciences and technology from the 1960s to the year 2000[J].Journal of Power Sources, 2001,(102):242-252. 被引量:1
  • 9Fadley C S, Wallace R A. Electrical conductivity of a polystyrene sulfonic acid membrane[J].Electropolymer studies II, 1968,115(12):1264-1270. 被引量:1
  • 10Pourcelly G, Oikonomou A, Gavach C. Influence of water content on the kinetics of counter-ion transport in perfluorosulphonic membranes[J].Journal of Electroanalytical Chemistry, 1990,(287):43-59. 被引量:1

共引文献20