Laser-induced breakdown spectroscopy was employed to determine the inorganic elements in coal. To improve the measurement's accuracy and precision, a new internal standardization scheme, which we named changed intern...Laser-induced breakdown spectroscopy was employed to determine the inorganic elements in coal. To improve the measurement's accuracy and precision, a new internal standardization scheme, which we named changed internal standardization, was compared with the traditional internal standardization and no internal standardization for the analysis of inorganic elements. The new internal standardization scheme used the atomic line of carbon at 247.86 nm and the molecular band of CN at 388.34 nm and C2 at 516.32 nm to normalize the lines of inorganic elements that were distributed in the same spectral channel. The performance of the utilization of the new internal standardization scheme was evaluated using a set of coal samples, including twenty calibration samples and five validation samples. The results show that the coefficients of determination R2 and the slope of calibration models coupled with changed internal standard- ization are better than that of the calibration models coupled with fixed internal standardization and no internal standardization. Moreover, the measurement accuracy and reproducibility of changed internal standardization for the analysis of five validation samples also yielded further improvement. The results that we obtained suggest that changed internal standardization could compensate for the matrix effects, as well as the influence of the difference in the spectral response of the light collection system.展开更多
基金supported by Open Research Fund of State Key Laboratory of Pulsed Power Laser Technology of China(No.SKL2013KF03)National Natural Science Foundation of China(Nos.51206055,51476061)+3 种基金the Fundamental Research Funds for the Central Universities of China(No.2014ZZ0014)the New Star of Pearl River on Science and Technology of Guangzhou,China(No.2014J2200054)the Key Laboratory of Efficient and Clean Energy Utilization of Guangdong Higher Education Institutes of China(No.KLB10004)Guangdong Province Key Laboratory of Efficient and Clean Energy Utilization,China(No.2013A061401005)
文摘Laser-induced breakdown spectroscopy was employed to determine the inorganic elements in coal. To improve the measurement's accuracy and precision, a new internal standardization scheme, which we named changed internal standardization, was compared with the traditional internal standardization and no internal standardization for the analysis of inorganic elements. The new internal standardization scheme used the atomic line of carbon at 247.86 nm and the molecular band of CN at 388.34 nm and C2 at 516.32 nm to normalize the lines of inorganic elements that were distributed in the same spectral channel. The performance of the utilization of the new internal standardization scheme was evaluated using a set of coal samples, including twenty calibration samples and five validation samples. The results show that the coefficients of determination R2 and the slope of calibration models coupled with changed internal standard- ization are better than that of the calibration models coupled with fixed internal standardization and no internal standardization. Moreover, the measurement accuracy and reproducibility of changed internal standardization for the analysis of five validation samples also yielded further improvement. The results that we obtained suggest that changed internal standardization could compensate for the matrix effects, as well as the influence of the difference in the spectral response of the light collection system.