期刊文献+

磁感应方式电导率测量基础研究 被引量:17

The Basic Research of the Measurement of Conductivity in a Magnetic Induction Way
下载PDF
导出
摘要 本文提出了用于磁感应方式电导率测量的单通道硬件系统的原理和结构,在此基础上,以NaCl溶液和琼脂为目标,进行了一系列试验,初步评估了场域的敏感性分布,证明了补偿线圈可以有效提高系统的相对灵敏度,试验结果与理论分析的比较证明了试验结果的可靠性。 This paper presents the principle and structure of the hardware used in the single-channel measurement of conductivity in a magnetic induction way. Based on the hardware, a series of experiments are made over saline solution targets and agar targets, the sensitivity distribution are evaluated preliminarily, we proved that a higher relative sensitivity can be obtained by using a back-off coil, reliable conclusions are drawn from the comparison between theoretical analysis and experiment results.
出处 《中国医学物理学杂志》 CSCD 2004年第3期182-185,181,共5页 Chinese Journal of Medical Physics
基金 全军医药卫生科研基金课题(01Z078)
关键词 非接触 磁感应 磁感应断层成像 电导率 contactless magnetic induction magnetic induction tomography conductivity
  • 相关文献

参考文献7

  • 1Johannes Netz, Ewald Fornert, Sabine Elaagemannf. Contactless impedance measurement by magnetic induction-a possible method for investigation of brain impedance[J]. Physiol. Meas., 1993,14:463~471. 被引量:1
  • 2Griffiths H, Stewart WR, Gough W. Magnetic induction tomography-A measuring system for biological tissues[J]. Ann N Y Acad Sci., 1999,873:335~345. 被引量:1
  • 3Griffiths H. Magnetic induction tomography [J]. measurement science and technology, 2001,12:1126~1131. 被引量:1
  • 4Morris A, Griffiths H, Gough W. A numerical model for magnetic induction tomographic measurements in biological tissues[J]. Physiol. Meas., 2001,22:113~119. 被引量:1
  • 5李世俊,秦明新,董秀珍,尤富生,史学涛,付峰,刘锐岗,李江.非接触磁感应脑阻抗断层成像系统激励源设计[J].第四军医大学学报,2003,24(2):176-178. 被引量:10
  • 6李世俊,秦明新,董秀珍,尤富生,史学涛,付峰.非接触磁感应脑阻抗断层成像系统设计[J].中国医学物理学杂志,2003,20(1):55-58. 被引量:21
  • 7S Watsonl, R J Williams, H Griffiths, W Gough, A Morris. Magnetic induction tomography: phase versus vector-voltmeter measurement techniques[J]. Physiol. Meas., 2003,24:555~564. 被引量:1

二级参考文献9

  • 1[1]Griffiths H , Stewart W R , Gough W . Magnetic induction tomography. A measuring system for biological tissues[J]. Ann N Y Acad Sci, 1999; 873: 335-345. 被引量:1
  • 2[2]Ren JL, Lin JM, Gao CF, Dianci Jiance (Electromagnetism Detecting )[M]. Beijing: Jixie Gongye Chubanshe ( Beijing: Machine and Industry Publishing House), 2000: 45-48. 被引量:1
  • 3[3]Scharfetter H, Lackner K, Rosell J. Magnetic induction tomography: Hardware for multi- frequency measurements in biological tissues [J]. Physiol Meas, 2001;22(1):1-16. 被引量:1
  • 4[4]Netz J, Fornert E, Haagem ann S. Contactless impedance measurement by magnetic indution-a possible method for investigation of brain impedance[J]. Physiol Meas,1993; 14: 463-471. 被引量:1
  • 5[5]Delisa JA , Gans BM.Rehabilitation Medicine Principles and practice.[M] New York : Third Edition Lippioncott-Raven Publishers, 1998:78-104. 被引量:1
  • 6[6]Korjenevsky A, Cherepenin V. Progress in realisation of magnetic induction tomography Ann [J]. New York Acad Sci, 1999; 873:346-352. 被引量:1
  • 7[7]Cao XW. Gaopin Dianlu Yuanli Yu Fenxi (The Principle and analyzing of High Frequency Circuit ) [M] . Xian: Dianzi Keji Daxue Chubanshe ( Xian: Xidian University publishing House) , 2001: 109-112. 被引量:1
  • 8[8]Miu HS. Dianliao Yu Guangliao ( Electrotherapeutics and phototherapy ) [M].Shanghai Kexue Jishu Chubanshe ( Shanghai : Science and Technology Publishing House ) , 1990 : 269-271. 被引量:1
  • 9董秀珍,秦明新,刘锐岗,汤孟兴,史学涛,尤富生,付峰.影响生物电阻抗断层成像质量的因素[J].中国医学物理学杂志,2001,18(4):206-207. 被引量:9