支抗种植体不同螺纹顶角对骨界面应力分布的影响 被引量:13

The Effects of Different Implant Screw Angle on Stress Distribution in Implant-bone Interface.
下载PDF
导出
摘要 目的 研究刃状螺纹型支抗种植体的螺纹顶角改变对骨界面应力分布的影响,以供临床筛选合适的支抗种植体。方法 用三维有限元方法给种植体施加150g近远中方向的载荷,分别对螺纹顶角为60°、75°、120°的三种支抗种植体-骨界面进行应力分析。结果 三种螺纹顶角种植体颈部的Von-Mises应力值分别为0.533 0 MPa、0.632 0 Mpa、0.591 0 Mpa;位移值分别为0.1630μm、0.1590μm、0.1520μm。结论 螺纹顶角为60°的种植体适合作正畸支抗体。 Objective To investigate the effects of different implant screw angle on stress distribution in implant-bone inter- face, providing the appropriate shape for orthodontic anchorage. Methods The stress and strain in implant-bone interface for im- plants with different screw angle(60°, 75°, 120°), which was 150 grams loaded mesiodistally, were analyzed by 3-dimesion finite element analysis method. Results The Von-mises stress at cervices of these three implant screw angles was 0. 533 0 MPa, 0. 632 0 MPa and 0. 591 0 MPa respectively. The alveolar bone deformation at the cervix of implants were 0. 163 0 μm, 0. 159 0 μm and 0. 152 0 μm respectively. Conclusion The implant with 60°screw angle was suitable for orthodontic anchorage.
出处 《广东牙病防治》 2003年第4期259-261,共3页 Journal of Dental Prevention and Treatment
基金 广东省医学科学技术研究基金(编号:A2002944)
关键词 支抗种植体 螺纹顶角 骨界面 应力分布 Implant anchorage Three-dimesion finite element analysis Anchorage implant-bone interface Screw angle
  • 引文网络
  • 相关文献

参考文献11

  • 1[1]Mailath G, Stoiber B, Atzek C, et al. Bone resorption at the entry of osseointegrated implants-a biomechanical phenomenon: Finite element study.Z Stomatol, 1989,86(4) :207-216 被引量:1
  • 2[2]Holmgren EP, Scekinger RJ, Kilgren LM, et al. Evaluating parameters of osseointegrated dental implants using finite element analysis-a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. J Oral Implantol, 1998,24(2):80-88 被引量:1
  • 3兰泽栋,林珠,龙红月,周磊.支抗种植体—骨界面三维有限元模型的建立[J].中国口腔种植学杂志,2003,8(2):57-60. 被引量:19
  • 4[4]Siegele D, Soltesz U. Numerical investigations of the influence of implant shape on stress distribution in the jaw bone. Int J Oral Maxillofac Implant,1989,4(4) :333-340 被引量:1
  • 5[5]Tsutsumi S, Fukuda S, Tani Y. Biomechanical designing of implant. J Dent Res, 1989,68(Special Issue) :766 被引量:1
  • 6[7]Gray JB, Steen ME, King GJ, et al. Studies on the efficacy of implants as orthodontic anchorage. Am J Orthod, 1983,83(4) :311-317 被引量:1
  • 7[9]Glatzmaier J, Wehrcein H, Diedrich P, et al. Biodegradable implants for orthodontic anchorage. Eur J Orthod, 1996,18(5) :465-469 被引量:1
  • 8[10]Hanawa T. Titanium and its oxide film: A substrate for formation of apatite. In: Davies JE (ed). Bone-bimaterial Interface. Toronto: Univ of Toronto Press. 1991.49 被引量:1
  • 9[11]GrayJB, StcenME, KingGJ, etal. Studies on the efficacy of implants as orthodontic anchorage. Am J Orthod, 1983,83 (4): 311-317 被引量:1
  • 10[12]HursonS. Thread implant design critera. Int J Dent Symp, 1994,2(1):38 被引量:1