摘要 日本CHINO公司生产的IR-RST系列标准辐射温度计由于没有设计温度显示装置,无法实时显示温度,针对此类型温度计设计了一款专用的电压测量和温度显示,同时可提供24 V直流电压的电测显示仪表。利用高等级的电压源和数字电压表,修正了本专用数字表的误差系数,并考核了稳定性,得到其最大相对误差不超过0. 03%。与一款名义波长为0. 65μm的IR-RST辐射温度计配合使用,通过TG HT-9500型高温炉对其组合进行内插分度,实验结果证实分度后校准准确度优于0. 03%,1300℃时系统测量结果的扩展不确定度为0. 62℃。 The IR-RST standard radiation thermometers manufactured by CHINO Company,Limited,Japan,which can not display the temperature in real-time because the display unit hadn’t been designed. Therefore,an electrical measurement display device was designed which can measure the voltage,display the temperature and supply 24 V DC voltage for this kind of infrared radiation thermometer. Calibrated a against high-level voltage power source and a voltmeter,the correction coefficients of the display instrument were got,the stabilitywas evaluated as well,the maximum relative error of voltage measurement can be achieved with in 0. 03%. Assembled with a 0. 65 μm IR-RST radiation thermometer,it was calibrated against with a TG HT-9500 type high-temperature furnace as the source,the accuracy was better than 0. 03%confirmed by the experimental results,and the extended uncertainty of the calibration system was 0. 62 ℃ at 1 300 ℃.
作者 盛山菊 卢小丰 唐小洁 宦克为 王雪莲 SHENG Shan-ju;LU Xiao-feng;TANG Xiao-jie;HUAN Ke-wei;WANG Xue-lian(Changchun University of Science&Technology,Changchun,Jilin 130022,China;National Institute of Metrology,Beijing 100029,China;Xi'an Polytechnic University,Xi'an,Shaanxi 710048,China)
出处 《计量学报》 CSCD 北大核心 2019年第1期20-24,共5页 Acta Metrologica Sinica
基金 科技部NQI专项(2017YFF0205902)
关键词 计量学 红外辐射温度计 电测显示仪表 校准 不确定度 metrology infrared radiation thermometer digital display instrument calibration uncertainty
  • 相关文献

参考文献7

二级参考文献51

共引文献42

同被引文献39

引证文献3

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部