摘要 目的:对多参数水质分析仪法与国标法测定水中游离余氯进行对比研究。方法:采用氯标准溶液对多参数水质分析仪的内置工作曲线进行验证,采用加标回收实验对多参数水质分析仪法进行测试。结果:多参数水质分析仪工作曲线准确可靠,回收率在80.0%~95.0%,精密度在0.38%~6.24%,符合标准要求。结论:与国标法相比,多参数水质分析仪法检测速度更快,精密度更高,适用于生活饮用水中游离余氯的快速检测。 Objective:To compare the determination of free residual chlorine in water by multi-parameter water quality analyzer method and national standard method.Method:The built-in working curve of the multi-parameter water quality analyzer was verified by chlorine standard solution,and the multi-parameter water quality analyzer method was tested by recovery test.Result:The working curve of the multi-parameter water quality analyzer was accurate and reliable,with a recovery rate of 80.0% to 95.0% and a precision of 0.38% to 6.24%,which met the standard requirements.Conclusion:Compared with the national standard method,the multi-parameter water quality analyzer method has faster detection speed and higher precision,and is suitable for the rapid detection of free residual chlorine in drinking water.
作者 于卓然 段瑞晶 薛钦 王丽静 杜利君 薛颖 刘春光 YU Zhuoran;DUAN Ruijing;XUE Qin;WANG Lijing;DU Lijun;XUE Ying;LIU Chunguang(Shanghai Customs,Shanghai 200000,China;Taiyuan Customs Technology Center,Taiyuan 030024,China;Shanxi Xiaolangdi Yellow River Diversion Water Group Co.,Ltd.,Yuncheng 044000,China)
出处 《现代食品》 2024年第19期200-204,共5页 Modern Food
关键词 多参数水质分析仪 国家标准方法 游离余氯 生活饮用水 muti-parameter water quality analyzer national standard method free residual chlorine drinking water
  • 相关文献

参考文献11

二级参考文献77

共引文献75

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部