期刊文献+

东芝Aquilion 16 CT磁电阻传感器故障维修1例

Fault repair of Toshiba Aquilion 16 CT MR sensor:1 case report
下载PDF
导出
摘要 阐述了东芝Aquilion系列螺旋CT相较于传统CT的优势,介绍了东芝Aquilion 16 CT使用过程中出现的1例磁电阻传感器故障,分析了故障产生的原因,提出了具体的解决方法,为医学工程人员维修类似故障提供了参考。 The advantages of Toshiba Aquilion series spiral CT were described over the traditional CT.One case of MR sensor fault of Toshiba Aquilion 16 CT and its cause was introduced,and the countermeasure was proposed accordingly.References were provided for medical engineers to treat similar faults.
作者 陶亚飞 王延辉 TAO Ya-fei;WANG Yan-hui(Department ofRadiography,the 980th Hospital of Joint Logistics Support Force,Shijiazhuang 050082,China;Department of Medical Engineering,the 980th Hospital of Joint Logistics Support Force,Shijiazhuang 050082,China)
出处 《医疗卫生装备》 CAS 2024年第2期118-120,共3页 Chinese Medical Equipment Journal
关键词 东芝Aquilion 16 CT 磁电阻传感器 故障维修 Toshiba Aquilion 16 CT MR sensor fault repair