期刊文献+

典型电流体动力学传导泵性能的实验研究

Experimental study of performance of typical electrohydrodynamic conduction pumping
下载PDF
导出
摘要 为考查典型电流体动力学传导泵性能,推动其应用于电子元器件微通道液冷,设计了平嵌电极与多孔电极电导泵,并搭建电导泵实验系统,测试了这2种电导泵泵送介电液体HFE-7100的性能。结果表明:施加10 kV直流电压时,6对平嵌电极电导泵产生流量194.5 mL/min,动压40.3 Pa,最大静压72.1 Pa;3对孔径为1.5 mm的多孔电极电导泵产生流量438.2 mL/min,动压173.0 Pa,最大静压363.5 Pa。可见,相比平嵌电极电导泵,多孔电极电导泵具有更好的性能。 In order to test the performance of typical electrohydrodynamic(EHD)conduction pumps and promote their applications in the microchannel liquid cooling of the electronic components,the flushed electrode and perforated electrode EHD conduction pumps are designed,and an experimental system of the EHD conduction pumping is established to test the performance of the two typical kinds of the EHD conduction pumps to transport the dielectric liquid HFE-7100.It is shown that at the applied voltage of 10 kV DC,the flow rate,the dynamic pressure,and the maximum static pressure of the EHD conduction pumping with six pairs of flushed electrodes are 194.5 mL/min,40.3 Pa,and 72.1 Pa,respectively;while the flow rate,the dynamic pressure,and the maximum static pressure for three pairs of the perforated electrode EHD conduction pumping with a diameter of 1.5 mm are 438.2 mL/min,173.0 Pa,and 363.5 Pa,respectively.So it can be said that the perforated electrode has better performance than the flushed electrode.
作者 杜中林 马诗远 易红亮 吴健 DU Zhonglin;MA Shiyuan;YI Hongliang;WU Jian(Key Laboratory of Aerospace Thermophysics of MIIT,School of Energy Science and Engineering,Harbin Institute of Technology,Harbin 150001,China)
出处 《航天器环境工程》 北大核心 2022年第6期614-619,共6页 Spacecraft Environment Engineering
基金 国家自然科学基金项目(编号:12172110) 北京卫星环境工程研究所创新基金项目(编号:CAST-BISEE 2019-012)。
关键词 电导泵 平嵌电极 多孔电极 介电液体 空间热管理 EHD conduction pumping flushed electrode perforated electrode dielectric liquid space thermal management
  • 相关文献

参考文献6

二级参考文献129

共引文献35

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部