摘要 目的:对2018-2020年使用BacT/ALERT 3D微生物培养系统进行无菌试验样本进行回顾性分析,特别是对假阳性和假阴性样本的具体分析,提高细菌检测能力。方法:按常规检测方法、仪器设备和试剂,对质控抽检样本、中国国际输血感染预防和控制(CITIC)样本进行检测,对检测结果进行分析和比较。结果:2018-2020年细菌检测共出现假阳性3例,假阴性6例,CITIC总体符合率为90.91%,阳性标本检出时间SDI优秀率为83.33%,假阳性原因为仪器使用方面、人为污染等,假阴性主要为低浓度细菌未检出。结论:保障血液安全非常重要,特别是输血不良反应的及时反馈,如果能够利用互联网模式,建立血站与医院之间的信息系统共享,将会有效提高采供血工作效率。 Objective:To analyze retrospectively the results of sterility test samples using BacT/ALERT 3D microbial culture system from 2018 to 2020,specifically analyze false positive sample and false negative samples,and improve the ability to detect bacteria.Methods:According to the routine testing methods,instruments and reagents,the quality control sampling samples and China International Transfusion Infection Control(CITIC)samples were tested,the test results were analyzed and compared.Results:There were 3 false positive cases and 6 false negative cases in bacterial detection from 2018 to 2020,the overall coincidence rate of CITIC was 90.91%,and the excellent rate of SDI in the positive sample detection time was 83.33%.The causes of false positive were the use of the instruments and human pollution,the causes of false negative were mainly that low concentration of bacteria was not detected.Conclusion:It is very important to ensure blood safety,especially the timely feedback of adverse reactions associated with blood transfusion,if the Internet mode can be used to establish information system sharing between blood stations and hospitals,it will effectively improve the efficiency of blood collection and supply.
作者 石圆圆 郁正芹 倪善红 Shi Yuan-yuan;Yu Zheng-qin;Ni Shan-hong(Lianyungang Red Cross Center Blood Station Quality Control Department,Lianyungang 222000,Jiangsu Province,China;Kangda College of Nanjing Medical University,Lianyungang 222000,Jiangsu Province,China)
出处 《中国社区医师》 2022年第32期164-166,共3页 Chinese Community Doctors
基金 南京医科大学康达学院科研发展基金课题(编号:KD2019KYJJZD003)。
关键词 细菌检测 回顾性分析 BacT/ALERT 3D微生物培养系统 室间质评 Bacterial detection Retrospective analysis BacT/ALERT 3D microbial culture system External quality assessment
  • 相关文献

参考文献4

二级参考文献25

  • 1陆英,陈运贤,林东军,刘赴平.全自动血液分析仪在检测血小板细菌污染中的应用[J].中国输血杂志,2006,19(6):459-460. 被引量:3
  • 2Breeher ME, Hay SN, Rothenberg SJ. Monitoring of apheresis platelet bacterial contamination with an automated liquid culture system : a university experience. Transfusion, 2003;43:974 -978 被引量:1
  • 3AuBuchon JP, Cooper LK, Leach MF, et al. Experience with universal bacterial culturing to detect contamination of apheresis platelet units in a hospital transfusion service. Transfusion, 2002 ;42:855 - 861 被引量:1
  • 4Brecher ME, Hay SN, Rothenberg SJ. Validation of BacT/ALERT plastic culture bottles for use in testing of whole-blood-derived leukoreduced platelet-rich-plasma-derived platelets. Transfusion, 2004; 44:1174 - 1178 被引量:1
  • 5国家技术监督局.献血者健康检查要求.国家标准GB18467-2001 被引量:1
  • 6Brecher ME, Hay SN, Rothenberg SJ. Evaluation of a new generation of plastic culture bottles with an automated microbial detection system for nine common contaminating organisms found in PLT components. Transfusion, 2004 ;44:359 - 363 被引量:1
  • 7McDonald CP, Rogers A, Cox M, et al. Evaluation of the 3D BacT/ ALERT automated culture system for the detection of microbial contamination of platelet concentrates, Transfus Meal, 2002 ; 12 : 303 - 309 被引量:1
  • 8Brecher ME, Holland PV, Pineda AA, et al. Growth of bacteria in inoculated platelets : implications for bacteria detection and the extension of platelet storage. Transfusion, 2000 ; 40 : 1308 - 1312 被引量:1
  • 9WS/T414-2013.室间质量评价结果应用指南. 被引量:2
  • 10国家药典委员会.中华人民共和国药典三部,2010:附录89-92. 被引量:1