7.0T MRI与3.0T MRI膝关节图像质量对比分析

Comparative analysis of knee image quality between 7.0T MRI and 3.0T MRI
下载PDF
导出
摘要 目的对比分析7.0T MRI和3.0T MRI两种不同场强的膝关节图像质量。方法使用3.0T MRI和7.0T MRI扫描仪,在相同时间内使用相同序列对受试者(n=16)进行膝关节成像。分别对肌肉、骨骼、韧带、软骨、神经的信噪比(SNR)和对比信噪比(CNR)进行定量分析,并由2位有丰富经验的放射科肌肉骨骼诊断医师对图像进行解剖结构可视化评估。结果7.0T MRI下膝关节图像的SNR和CNR明显优于3.0T MRI,差异具有统计学意义(P<0.05)。7.0T MRI膝关节图像具有更高的清晰度和对比度。结论相同时间内,7.0T MRI扫描的膝关节图像SNR和CNR更高,能够更清晰地显示膝关节的解剖结构,有助于提高诊断的准确性。 Objective To compare and analyze the knee image quality of two different magnetic field intensity of 7.0T MRI and 3.0T MRI.Methods Knee joint imaging was performed on subjects(n=16)with 3.0T MRI and 7.0T MRI scanner at the same time using the same sequence.The signal to noise ratio(SNR)and contrast signal to noise ratio(CNR)of muscle,bone,ligament,cartilage and nerve were quantitatively analyzed,and the anatomical structure of the images were visualized and evaluated by two experienced musculoskeletal diagnostic radiologists.Results The SNR and CNR of the knee images under the 7.0T MRI were significantly better than those under 3.0T MRI,and the difference was statistically significant(P<0.05).The knee images of the 7.0T MRI have higher clarity and contrast.Conclusion Within the same time,the SNR and CNR of the knee images scanned by 7.0T MRI are higher,which can show the anatomical structure of the knee more clearly,and help to improve the accuracy of diagnosis.
作者 李春霖 秦诗逸 甄志铭 陈家飞 LI Chun-lin;QIN Shi-yi;ZHEN Zhi-ming;CHEN Jia-fei(Department of Radiology,First Affiliated Hospital of Army Medical University,Chongqing 400038,China;Department of Vascular Surgery,First Affiliated Hospital of Army Medical University,Chongqing 400038,China)
出处 《局解手术学杂志》 2022年第10期886-889,共4页 Journal of Regional Anatomy and Operative Surgery
基金 重庆市临床重点专科精品项目(CQZDZK007)。
关键词 7.0T MRI 膝关节 图像质量 信噪比 对比信噪比 7.0T MRI knee joint image quality signal to noise ratio contrast signal to noise ratio
  • 相关文献

参考文献12

二级参考文献85

  • 1倪萍,蔡华.磁共振成像设备新进展[J].医疗设备信息,2005,20(12):1-4. 被引量:9
  • 2朱弋,王斌,杨舒波,李晶.影响核磁共振图像均匀度的因素和检定方法[J].医疗装备,2006,19(7):16-17. 被引量:8
  • 3倪萍,陈自谦,肖慧,钱根年,陈景华.磁共振螺旋桨技术与图像质量的相关性研究[J].医疗设备信息,2007,22(6):78-80. 被引量:9
  • 4赵海涛.磁共振成像原理及应用[M].西安:第四军医大学出版社,2004.157-170. 被引量:2
  • 5Ladd ME. High-field-strength magnetic resonance: potential and limits[J].Top Magn Reson Imaging,2007,18:139-152. 被引量:1
  • 6Ethofer T,Mader I,Seeger U,et al.Comparison of longitudinal metabolite relaxation times in different regions of the human brains at 1.5 and 3 Tesla[J].Magn Reson Med,2003,50:1296-1301. 被引量:1
  • 7MansfieldP MorrisPG.生物医学中的核磁共振成像[M].杭州:浙江大学出版社,1987.. 被引量:3
  • 8Price RR, Axel L, Morgan T, et al. Quality assurance methods and phantoms for magnetic resonance imaging. Report of AAPM Nuclear Magnetic. Resonance Task Group. No.1. Med Phys, 1990,17:287-295. 被引量:1
  • 9Och JG, Clarke GD, Sobol WT, et al. Acceptance testing of magnetic resonance imaging systems. Report of AAPM Nuclear Magnetic Resonance Task Group. No.6. Med Phys, 1992,19:217-229. 被引量:1
  • 10Kaufman L, Kramer DM, Crooks LE, et al. Measuring signal-to-noise ratios in MR imaging. Radiology, 1989,173:265-267. 被引量:1

共引文献646

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部