期刊文献+

电感耦合等离子体发射光谱法测定含硼水中银离子

Determination of Silver Ions in Boron Water by Inductively Coupled Plasma Optical Emission Spectrometry
下载PDF
导出
摘要 研究并建立了电感耦合等离子体发射光谱法测定含硼水中银离子的分析方法,通过试验选定仪器工作条件,采用硼基体匹配消除硼对银离子的分析干扰,该方法的测定下限为3μg/L,回收率在90%~110%,该方法具有较高的准确度,满足含硼水中银离子的测定要求。 An analytical method for the determination of silver ions in boron water by inductively coupled plasma optical emission spectrometry was studied and established in this paper.Through the experiments,the working conditions of the instrument were selected,and the boron matrix matching was used to eliminate the analytical interference of boron to silver ions.By this method,the lower limit of determination is 3μg/L,and the recovery is between 90% and 110%,indicating that the method has high accuracy and meets the requirements for determination of silver ion in boron water.
作者 韩剑 程建华 Han Jian;Cheng Jian-hua
出处 《化工设计通讯》 CAS 2022年第3期96-98,共3页 Chemical Engineering Design Communications
基金 国家自然科学基金项目(51904192,52074082,51804071) 辽宁省博士科研启动基金计划项目(2020-BS-155) 辽宁省教育厅基本科研项目(面上项目)(LJKZ0248) 沈阳理工大学引进高层次人才科研支持计划项目(1050002000611)。
关键词 电感耦合等离子体发射光谱法 银离子 准确度 inductively coupled plasma emission spectrometry silver ion accuracy