全自动石墨消解-连续光源火焰原子吸收测定土壤中的铅 被引量:4

Determination of Lead in Soil by Automatic Graphite Digestion and Continuum Source Atomic Absorption Spectrometer
下载PDF
导出
摘要 采用全自动石墨消解前处理方法,建立连续光源火焰原子吸收法测定土壤中铅元素检测方法。试验表明,铅有较好的线性,相关系数达0.9998;方法检出限为0.5 mg/kg,测定下限为2.0 mg/kg;相对标准偏差在0.6%~3.4%之间,相对误差在1.0%~2.2%之间。该方法操作简便快速、灵敏度高、准确性好。 A new rapid method was established to determine lead in soil by automatic graphite digestion and continuum source atomic absorption spectrometer.The test shows,lead had a good linearity,the correlation coefficient was 0.9998.The method detection limit was 0.5 mg/kg,lower detection limit was 2.0 mg/kg.The relative standard deviation was 0.6%to 3.4%,the relative error was 1.0%to 2.0%.The method was easy and quick,had high sensitivity and accuracy.
作者 吴晶 毛慧 朱小梅 Wu Jing;Mao Hui;Zhu Xiaomei(Jiangsu Taizhou Environmental Monitoring Center,Taizhou 225300,China)
出处 《山东化工》 CAS 2021年第21期92-93,共2页 Shandong Chemical Industry
关键词 连续光源 原子吸收 全自动石墨消解 土壤 continuum source atomic absorption automatic graphite digestion soil