期刊文献+

X射线环境辐射监测仪器校准方法的研究 被引量:10

Research on Calibration Method of X-ray Environmental Radiation Monitoring Instrument
下载PDF
导出
摘要 环境辐射监测仪表作为微弱放射性监测计量器具,它的量值溯源是一个急需解决的问题。由于环境辐射剂量率低且电离信号微弱,故通常采用大体积电离室或者充压电离室进行测量。依托60~250 k V X射线空气比释动能基准电离室,在完成重过滤窄谱X射线辐射质空气比释动能绝对测量的基础上,通过逐级替代法完成大体积环境辐射监测仪器的校准。测量结果不确定度为5.6%(k=2),实现环境水平X射线空气比释动能测量量值溯源,为国内环境辐射监测仪器在低剂量率水平的性能评价提供计量保障。 As a measuring instrument for weak radioactivity monitoring, the traceability of the environmental monitoring is an urgent problem to be solved. Because the environmental radiation dose rate is low and the ionization signal is weak,it is usually measured by large volume ionization chamber or charging pressure ionization chamber. On the based of primary standards of 60 ~ 250 k V X-ray air kerma and absolute measurement of the air kerma under the narrow spectrum series X-ray radiation quality,radiation environmental monitoring instruments were calibratied by successive substitution method. The uncertainty of measurement measurement result is better than 5. 6%( k = 2). It realize the traceability of environmental level X-ray air kerma measurement,and provide metrological support for the performance evaluation of environmental radiation monitoring instruments at low dose rate level.
作者 宋飞 赵瑞 丁卫撑 张德亮 杨扬 吴金杰 SONG Fei;ZHAO Rui;DING Wei-cheng;ZHANG De-liang;YANG Yang;WU Jin-jie(College of Nuclear Technology and Automation Engineering,Chengdu University of Technology,Chengdu,Sichuan 610059,China;National Institute of Metrology,Beijing 100029,China)
出处 《计量学报》 CSCD 北大核心 2021年第9期1237-1243,共7页 Acta Metrologica Sinica
基金 国家重点研发计划(2017YFF0205101)。
关键词 计量学 X射线 空气比释动能 逐级替代法 溯源 metrology X-ray air kerma standard successive substitution method traceability
  • 相关文献

参考文献16

二级参考文献90

共引文献57

同被引文献74

引证文献10

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部