摘要 通过分析Brilliance64排螺旋CT的工作原理、正常工作干扰因素及常见故障案例,为排除CT设备在工作中遇到的故障提供解决方案。明确CT设备工作原理与各种干扰因素和故障排除方法后,使CT机得到有效保养与维护,及时排除运行中发生的故障,为设备正常工作提供有效保障。 To solve the problems encountered in the work of CT equipment by analyzing the working principle,interference factors of normal work and common troubleshooting measures of Philips Brilliance 64-slice spiral CT.After clarifying the working principle of CT equipment,various interference factors and troubleshooting methods,the CT machine can be effectively maintained,which can quickly solve the faults that occurred during operation and provide effective guarantees for the normal operation of the equipment.
作者 王惟 张勇 周航旭 WANG Wei;ZHANG Yong;ZHOU Hang-xu(State-owned Assets Management Office,The First Affiliated Hospital of Xi'an Jiaotong University,Xi'an 710061,China;不详)
出处 《中国医学装备》 2021年第3期190-192,共3页 China Medical Equipment
关键词 计算机断层扫描 工作原理分析 干扰因素 故障排除 维护保养 Computed tomography(CT) Working principle analysis Interference factor Fault elimination Maintenance