摘要
目的:分析主动脉内经皮左心室辅助泵出口处的径向血流及其对主动脉内皮细胞造成损伤的可能性。方法:运用计算流体力学(CFD)对一款经皮左心室辅助泵在主动脉直径分别为20、30、40 mm 3种情况下进行非定常数值仿真,分析血泵出口处的流场分布。通过研究其速度、压力和剪切应力分布情况来分析血泵径向血流对主动脉血管内皮细胞的影响。结果:数据结果显示,血泵出口处的径向血流对3个模型额外增加的正应力分别为24、17、8 mmHg,近壁面剪切应力大于25 Pa的比例分别为19.3%、13.6%、3.0%。结论:实验表明,经皮左心室辅助泵的植入会增大主动脉压和近壁面剪切应力,而且主动脉直径越小,其效果越明显,因此对于主动脉直径偏小、患有高血压或动脉病变的患者应谨慎使用。
Objective To analyze the radial flow at the outlet of intra-aorta percutaneous left ventricular assist pump and itspossibility of damage to aortic endothelial cells.Methods Based on computational fluid dynamics(CFD),the numerical imitation of unsteady flow was carried in a percutaneous left ventricular assist pump with different aortic diameters of 20,30 mm,and 40 mm,and the flow field distribution at the outlet of the pump was analyzed.The effects of radial blood flow on aortic vascular endothelial cells were analyzed by studying the distribution of velocity,pressure and shear stress.Results The results showed that the radial blood flow at the outlet of the pump increased the normal stresses of the three models by respectively 24,17 mmHg,and 8 mmHg,and the proportions of near-wall shear stress greater than 25 Pa are 19.3%,13.6%,and 3%,respectively.Conclusions Experiments showed that the percutaneous left ventricular assist pump can increase the aortic pressure and near-wall shear stress,and the smaller the aortic diameter,the more obvious the effect.Therefore,patients with small diameter of aorta,hypertension or arterial lesions should be cautious to use.
作者
王飞
程云章
郑淇文
张金丽
唐洁
WANG Fei;CHENG Yunzhang;ZHENG Qiwen;ZHANG Jinli;TANG Jie(School of Medical instrument and Food Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;ForQaly Medical(Shanghai)Company Limited,Shanghai 201138,China)
出处
《中国医学物理学杂志》
CSCD
2020年第12期1573-1578,共6页
Chinese Journal of Medical Physics
基金
上海工程技术研究中心资助项目(18DZ2250900)。
关键词
经皮左心室辅助泵
数值模拟
剪切应力
压力
主动脉
Percutaneous left ventricular assist pump
numerical simulation
shear stress
pressure
aorta