摘要
目的探讨比较常规乳腺X线摄影(MG)与对比增强乳腺能谱摄影(CESM)对乳腺良恶性疾病的诊断价值。方法对2018年5月~2020年4月同时完成CESM和MG检查并最终获得病理诊断的74例患者纳入研究,均为女性,年龄22~56岁(41.02±8.24岁)。采用Stenographe Essential全数字化乳腺机,经由上肢静脉注射碘对比剂后采用头尾位(CC位)和内外斜位(MLO位),一次压迫高低能曝光,得到低能图像和经过特定算法处理的高能减去低能的减影图像,由3名放射科乳腺专业医师对所得图像进行质量分析并做出影像学诊断,以病理为标准,比较CESM与MG对乳腺良恶性疾病诊断差异,并进一步对比两种方法诊断为同一肿瘤分级的诊断符合率差异,分析比较两种方法的诊断效。结果共40例乳腺恶性肿瘤、42例良性病变。MG在诊断良恶性病变的灵敏度、特异度、阳性预测值、阴性预测值及准确率分别为77.50%、80.95%、79.48%、70.27%及79.07%,而CESM的灵敏度、特异度、阳性预测值、阴性预测值及准确率分别为100%、90.47%、90.90%、100%及95.12%(P<0.05)。MG在诊断乳腺肿瘤分级3~5级中的诊断符合率分别为79.00%、66.70%、86.70%、85.70%及100%,CESM为100%、92.30%、87.50%、90%及100%(P<0.05)。结论CESM诊断乳腺良恶性疾病优于MG,尤其是在诊断3级及4A级的肿瘤中。
Objective To investigate the diagnostic value of comparing conventional mammography(MG)and contrastenhanced spectralmammography(CESM)for benign and malignantbreast diseases,using pathological results as the gold standard.Methods Seventy-four patients who completed CESM and MG tests and obtained pathological diagnosis from May 2018 to April 2020 were included in the study.All patients were female,aged 22-56 years,with an average age of 41.02±8.24 years old.Using the American GE Stenographe Essential fully digital mammary gland machine,through the upper limb after intravenous injection of iodine contrast agent USES the end(CC)and internal and external oblique(MLO),by a pressure high and low to exposure,we obtain low energy image and subtraction angiography image of high energy subtracting low energy by special algorithm.Three radiology breast specialist performed quality analysis of the images and made imaging diagnosis.Taking pathology as the standard,the difference in the diagnosis of benign and malignant breast diseases between cesm and Mg was compared,and the diagnostic coincidence rate of the two methods for the same tumor grade was further compared,and the diagnostic efficacy of the two methods was analyzed and compared.Results The sensitivity,specificity,positive predictive value,negative predictive value and accuracy of MG in the diagnosis of benign and malignant breast lesions were 77.50%,80.95%,79.48%,70.27%and 79.07%,respectively.The sensitivity,specificity,positive predictive value,negative predictive value and accuracy of CESM were 100%,90.47%,90.90%,100%and 95.12%respectively(P<0.05).In the study,the diagnostic coincidence rate of MG in grade 3-5 breast tumor grade was 79.00%,66.70%,86.70%,85.70%and 100%respectively,while CESM was 100%,92.30%,87.50%,90%and 100%(P<0.05).Conclusion CESM is superior to conventional mammography in the diagnosis of benign and malignant breast diseases,especially in the diagnosis of grade 3and 4Atumors.
作者
邱子维
高文
张立
谢艳碧
陈静
文戈
QIU Ziwei;GAO Wen;ZHANG Li;XIE Yanbi;CHEN Jing;WEN Ge(Department of Radiology,Nanfang Hospital,Southern Medical University,Guangzhou 510515,China;Department of Radiology,Guangzhou Development Zone Hospital,Guangzhou 510730,China)
出处
《分子影像学杂志》
2020年第4期610-614,共5页
Journal of Molecular Imaging
基金
广东省基础与应用基础研究基金(2020A1515010469)。
关键词
对比增强乳腺能谱摄影
乳腺摄影
乳腺良恶性疾病
contrast-enhanced spectralmammography
mammography
benign and malignant breast diseases