摘要
目的对多层共挤输液用膜拉伸强度测量的不确定度进行评定。方法通过建立测量不确定度的数学模型,分析重复性测量、拉力机拉力误差、厚度测量、宽度测量等因素的测量不确定度,并对其进行评定。结果拉伸强度的测量不确定度为(32.66±0.14)MPa,其中扩展不确定度U=0.14 MPa。结论重复性测量对于不确定度结果的影响最大。
Objective To analyze the factors affecting the uncertainty in measurement of tensile properties of multi-layer co-extrusion film for infusion.Methods Establishing the measuring model,to analyze and evaluate the effects of the main factors,such as the repeatability measurement,thickness and width measurement,equipment accuracy.Results The uncertainty in measurement of tensile properties is(32.66±0.14)MPa,and the expanded uncertainty is 0.14 MPa.Conclusion Repeatability is the main factor for uncertainty.
作者
邵磊
薛文静
张潇予
张强
SHAO Lei;XUE Wenjing;ZHANG Xiaoyu;ZHAGN Qiang(Qingdao Institute for Food and Drug Control,Qingdao 266071,China)
出处
《药学研究》
CAS
2020年第10期572-574,共3页
Journal of Pharmaceutical Research
关键词
多层共挤输液用膜
不确定度
拉伸强度
Multi-layer co-extrusion film for infusion
Uncertainty
Tensile properties