宫颈癌患者的膀胱充盈度一致性对放疗摆位误差的影响 被引量:19

Effect of consistency of bladder filling volume on set-up errors in radiotherapy for the patients with cervical cancer
原文传递
导出
摘要 目的研究宫颈癌患者的膀胱充盈度一致性对于精准放疗摆位的重要性。方法选取宫颈癌患者30例,随机分为观察组和对照组;观察组在CT定位前,利用膀胱容量测定仪测定膀胱充盈度,每次治疗时膀胱容量达到定位容量时再放疗;对照组自主憋尿后行定位及放疗。每周1次CBCT扫描,分别测得患者左右、背腹和头脚方向的摆位误差。结果观察组和对照组在左右、背腹、头脚三个方向上摆位误差分别为(1.71±0.48)mm、(1.43±0.68)mm、(1.97±0.52)mm和(1.79±0.54)mm、(1.61±0.80)mm、(3.10±1.01)mm。观察组头脚方向摆位误差明显小于对照组(t=–3.837,P<0.05);通过摆位误差计算出的外扩边界分别为4.611 mm、4.051 mm、5.289 mm和4.853 mm、4.585 mm、8.457 mm,观察组在头脚方向明显优于对照组。结论宫颈癌患者的膀胱充盈度一致性有利于减少放疗时的摆位误差,减少外扩和正常组织受照,因此有助于精准治疗。 Objective To study the importance of the consistency of bladder filling volume to the accuracy of radiotherapy positioning for patients with cervical cancer.Methods 30 patients with cervical cancer were randomly divided into two groups,the observation group and the control group.Before CT localization,the bladder filling volume of patients in the observation group was measured by bladder scanner.Radiotherapy was performed when the bladder volume was reached the fixed volume.Patients in the control group received radiotherapy immediately after spontaneous bladder control.CBCT scan was performed to measure set-up errors on directions of left/right,dorsal/ventral,and head/foot,respectively.Results In the three directions of left/right,dorsal/ventral,and head/foot,the set-up errors of the observation group and control group were(1.71±0.48)mm,(1.43±0.68)mm,(1.97±0.52)mm and(1.79±0.54)mm,(1.61±0.80)mm,(3.10±1.01)mm.The set-up errors of head/foot direction of patients in the observation group were significantly smaller than that of patients in the control group(t=−3.837,P<0.05).The outspread boundary calculated by positioning errors were 4.611 mm、4.051 mm、5.289 mm and 4.853 mm、4.585 mm、8.457 mm,respectively.The head and foot direction of the observation group is significantly better than that of the control group.Conclusion The consistency of bladder filling volume of patients with cervical cancer is good to improve radiotherapy positioning accuracy,reduce the external expansion and exposure of normal tissue,which can help accurate radiotherapy.
作者 李钰 高岩 刘世龙 董丽华 夏诚诚 LI Yu;GAO Yan;LIU Shilong;DONG Lihua;XIA Chengcheng(Department of Radiotherapy,First Hospital of Jilin University,Changchun 130021 China)
出处 《中国辐射卫生》 2020年第3期305-308,共4页 Chinese Journal of Radiological Health
关键词 宫颈癌 放射治疗 膀胱充盈度 摆位误差 Cervical Cancer Radiotherapy Bladder Fullness Set-up Errors
  • 相关文献

参考文献7

二级参考文献65

  • 1廖利民.神经泌尿学现状与前景[J].中华泌尿外科杂志,2004,25(9):646-648. 被引量:32
  • 2高丽娟,鞠彦合,刘丽岩,董向竹,李秀华,廖利民.膀胱容量测定仪在神经源性膀胱患者间歇导尿中的临床应用[J].中国康复理论与实践,2007,13(7):619-620. 被引量:15
  • 3Mell LK, Roeske JC, Mundt AJ. A survey of intensity modulated radiation therapy use in the United States. Cancer,2003,98:204- 211. 被引量:1
  • 4Roeske JC, Lujan A, Rotmensch J, et al. Intensity-modulated whole pelvic radiation therapy in patients with gynecologic malignancies. Int J Radiat Oncol Biol Phys ,2000 ,48 :1613-1621. 被引量:1
  • 5Mell LK, Tiryaki H, Ahn KH, et al. Dosimetric comparison of bone marrow-sparing intensity-modulated radiotherapy versus conventional techniques for treatment of cervical cancer. Int J Radiat Oncol Biol Phys,2008,71:1504-1510. 被引量:1
  • 6Chen MF, Tseng C J, Tseng CC, et al. Clinical outcome in post hysterectomy cervical cancer patients treated with concurrent cisplatin and intensity-modulated pelvic radiotherapy: comparison with conventional radiotherapy. Int J Radiat Oncol Biol Phys, 2007,67 : 1438-1444. 被引量:1
  • 7Chan P, Dinniwell R, Haider MA, et al. Inter-and intrafractional tumor and organ movement in patients with cervical cancer undergoing radiotherapy: a cinematic-MRI point-of-interest study. Int J Radiat Oncol Biol Phys ,2008,70 : 1507-1515. 被引量:1
  • 8van de Bunt L, van der Heide UA, Ketelaars M, et al. Conventional, eonformal and Intensity-modulated radiation therapy treatment planning of external beam radiotherapy for cervical cancer:the impact of tumor regression. Int J Radiat Oncol Biol Phys, 2006,64 : 189-196. 被引量:1
  • 9Jhingran A, Salehpour M, Sam M, et al. Vaginal motion and bladder and rectal volumes during pelvic intensity-modulated radiation therapy after hysterectomy. Int J Radiat Oncol Biol Phys, 2011, In Press. 被引量:1
  • 10Raaymakers BW, Lagendijk JJ, Overweg J, et al. Integrating a 1.5 TMRI scanner with a 6 MV accelerator:proof of concept. Phys Med Biol,2009,54:229-237. 被引量:1

共引文献93

同被引文献158

引证文献19

二级引证文献49

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部