新华Super6000型清洗消毒器工作原理和故障维修 被引量:2

Working Principle and the Troubleshooting of SHINVA Super6000 Cleaning and Disinfecting Apparatus
下载PDF
导出
摘要 本文首先对新华Super6000型清洗消毒器的结构和工作原理进行分析,并以该设备的报错“前门电机保护”“水未准备好”“设备静止和运行过程中均出现漏气”现象3类故障和维修过程为例,详细介绍相关故障的测试方法和故障排除过程,通过更换PLC连接线、气动阀、前门气缸后使设备恢复正常运行。期望本文能为故障的快速修复和定期巡查制度提供参考,使设备运行更加稳定,更好的为临床服务。 Firstly,this paper analyzed the structure and working principle of Xinhua Super6000 cleaning sterilizer.Then,taking three kinds of faults of the equipment,such as“front door motor protection”“water not ready”“air leakage in the process of equipment static and operation”,and the maintenance process as examples,the test method and troubleshooting process of relevant faults were introduced in detail.The equipment was restored to normal operation by replacing the PLC connecting line,pneumatic valve and front door cylinder.It is expected that this paper can provide a reference for the rapid repair of faults and regular inspection system,so that the equipment operation is more stable and better for clinical services.
作者 裘科烽 郑焜 郑彩仙 QIU Kefeng;ZHENG Kun;ZHENG Caixian(Department of Medical Equipment,The Children’s Hospital of Zhejiang University School of Medicine,National Clinical Research Center for Child Health,Hangzhou Zhejiang 310003,China)
出处 《中国医疗设备》 2020年第8期171-173,184,共4页 China Medical Devices
关键词 清洗消毒器 气动阀故障 行程开关 前门气缸 cleaning sterilizer pneumatic valve failure travel switch front door cylinder