期刊文献+

口内三维扫描技术临床应用精度的研究进展 被引量:12

Research and development of clinical application accuracy of intraoral three-dimensional scanning technology
原文传递
导出
摘要 口内三维扫描技术发展迅速,精度不断提高,已被广泛应用于口腔医学领域。在单冠修复、三单位固定桥以及单枚种植体上部修复方面,口内三维扫描仪的扫描精度可满足临床要求;但在跨度过大的固定修复领域,其扫描精度仍存在不足。由于光学扫描原理的局限,目前临床难以通过口内三维扫描仪直接获取功能整塑后的软组织形态,口内三维扫描技术尚存在技术瓶颈。本文通过回顾和总结常见口内三维扫描仪对口腔软硬组织的扫描精度,分析口内三维扫描技术的临床适用性及局限性,以期为临床决策提供参考。 Intraoral three-dimensional scanning technology has been adopted in stomatology field and developed rapidly with improved accuracy.At present,the scanning accuracy of the intraoral three-dimensional scanners used in restoration of single crown,three unit fixed partial bridge and single implant crown satisfies the clinical requirements.While for restoration of multi-unit fixed,the scanning accuracy is not enough.There are still technical bottlenecks for scanning functional motion boundary of soft tissue,especially for direct acquisition of edentulous functional impressions.Through literature review and analysis,we summarize the intraoral scanning accuracy of oral hard and soft tissues,and analyze the clinical applicability and limitations of intraoral scanning technology,providing reference for clinical application.
作者 曹悦 陈俊锴 赵一姣 王勇 孙玉春 Cao Yue;Chen Junkai;Zhao Yijiao;Wang Yong;Sun Yuchun(Center of Digital Dentistry,Faculty of Prosthodontics,Peking University School and Hospital of Stomatology&National Clinical Research Center for Oral Diseases&National Engineering Laboratory for Digital and Material Technology of Stomatology&Research Center of Engineering and Technology for Digital Dentistry of Ministry of Health&Beijing Key Laboratory of Digital Stomatology,Beijing 100081,China)
出处 《中华口腔医学杂志》 CAS CSCD 北大核心 2020年第3期201-205,共5页 Chinese Journal of Stomatology
基金 装备预研教育部联合基金(6141A0202604) 首都科技领军人才培养工程(Z191100006119022) 首都卫生发展科研专项(首发2018-2-4103)。
关键词 牙模型 精度 数字化印模技术 口内三维扫描技术 Dental models Accuracy Dental impression technique Intraoral three-dimensional scanning technique
  • 相关文献

参考文献2

二级参考文献22

  • 1韩景芸,孙玉春.全口义齿数字化设计系统[J].北京工业大学学报,2009,35(5):587-591. 被引量:10
  • 2孙玉春,吕培军,王勇,韩景芸,赵建江.计算机辅助设计与快速成形技术辅助制作全口义齿的探讨[J].中华口腔医学杂志,2007,42(6):324-329. 被引量:21
  • 3孙玉春,吕培军,王勇.参数化定位全口义齿人工牙三维图形数据库的研究[J].口腔颌面修复学杂志,2007,8(4):257-260. 被引量:12
  • 4Boeddinghaus M, Breloer ES. Rehmann P. et al. Accuracy of single-tooth restorations based on intraoral digilal and conventional impressions in patients[J]. Clin Oral lnvestig, 2015, 19(8): 2027-2034. DOI: 10.1007/s00784-015-1430-7. 被引量:1
  • 5Patzelt SB, Vonau S, Stampf S, et al. Assessing the feasibility and accuracy of digitizing edentulous jaws[J]. J Am Dent Assoe, 2013. 144(8): 914-920. 被引量:1
  • 6Nedelcu RG, Persson AS. Scanning accuracy and precision in 4 intraoral scanners: an in vitro comparison based on 3-dimensional analysis[J]. J Prosthet Dent, 2014, 112(6): 1461-1471. DOI: 10.1016/j.prosdent.2014.05.027. 被引量:1
  • 7Ender A, Mehl A. Aeeuracy of complete-arch dental impressions: a new method of measuring trueness and precision[J]. J Prosthet Dent, 2013, 109(2): 121-128. DOI: 10. I 016/S0022-3913( 13)60028-1. 被引量:1
  • 8中华人民共和冈国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T6379.1-2004,测量方法与结果的准确度(正确度与精密度)第1部分:总则与定义[S].北京:中国标准出版社,1994. 被引量:1
  • 9Ender A, Mehl A. Full arch scans: conventional versus digital impressions: an in-vitro study[J]. Int J Comput Dent, 2011, 14 (1): 11-21. 被引量:1
  • 10Li H, Lyu P, Sun Y, et al. A quantitative study of 3D-scanning frequency and Ad of tracking points on the tooth surface[J]. Sci Rep, 2015, 5: 14350. DOh 10.1038/srep14350. 被引量:1

共引文献42

同被引文献130

引证文献12

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部