A low-noise, high-SNR balanced homodyne detector for the bright squeezed state measurement in 1–100 kHz range 被引量:7

A low-noise, high-SNR balanced homodyne detector for the bright squeezed state measurement in 1–100 kHz range
下载PDF
导出
摘要 We report a low-noise, high-signal-to-noise-ratio(SNR) balanced homodyne detector based on the standard transimpedance amplifier circuit and the inductance and capacitance combination for the measurement of the bright squeezed state in the range from 1 kHz to 100 kHz. A capacitance is mounted at the input end of the AC branch to prevent the DC photocurrent from entering the AC branch and avoid AC branch saturation. By adding a switch at the DC branch,the DC branch can be flexibly turned on and off on different occasions. When the switch is on, the DC output provides a monitor signal for laser beam alignment. When the switch is off, the electronic noise of the AC branch is greatly reduced at audio-frequency band due to immunity to the impedance of the DC branch, hence the SNR of the AC branch is significantly improved. As a result, the electronic noise of the AC branch is close to -125 dBm, and the maximum SNR of the AC branch is 48 dB with the incident power of 8 mW in the range from 1 kHz to 100 kHz. The developed photodetector paves a path for measuring the bright squeezed state at audio-frequency band. We report a low-noise, high-signal-to-noise-ratio(SNR) balanced homodyne detector based on the standard transimpedance amplifier circuit and the inductance and capacitance combination for the measurement of the bright squeezed state in the range from 1 kHz to 100 kHz. A capacitance is mounted at the input end of the AC branch to prevent the DC photocurrent from entering the AC branch and avoid AC branch saturation. By adding a switch at the DC branch,the DC branch can be flexibly turned on and off on different occasions. When the switch is on, the DC output provides a monitor signal for laser beam alignment. When the switch is off, the electronic noise of the AC branch is greatly reduced at audio-frequency band due to immunity to the impedance of the DC branch, hence the SNR of the AC branch is significantly improved. As a result, the electronic noise of the AC branch is close to -125 dBm, and the maximum SNR of the AC branch is 48 dB with the incident power of 8 mW in the range from 1 kHz to 100 kHz. The developed photodetector paves a path for measuring the bright squeezed state at audio-frequency band.
作者 Jin-Rong Wang Qing-Wei Wang Long Tian Jing Su Yao-Hui Zheng 王锦荣;王庆伟;田龙;苏静;郑耀辉(State Key Laboratory of Quantum Optics and Quantum Optics Devices,Institute of Opto-Electronics,Shanxi University,Taiyuan 030006,China;Collaborative Innovation Center of Extreme Optics,Shanxi University,Taiyuan 030006,China)
出处 《Chinese Physics B》 SCIE EI CAS CSCD 2020年第3期215-220,共6页 中国物理B(英文版)
基金 Project supported by the National Natural Science Foundation of China(Grant Nos.11654002,61575114,11874250,and 11804207) the National Key Research and Development Program of China(Grant No.2016YFA0301401) the Program for Sanjin Scholar of Shanxi Province,China the Program for Outstanding Innovative Teams of Higher Learning Institutions of Shanxi,China the Fund for Shanxi“1331 Project”Key Subjects Construction,China Key Research and Development Projects of Shanxi Province,China(Grant No.201903D111001)。
关键词 quantum optics PHOTODETECTOR LOW-NOISE audio band quantum optics photodetector low-noise audio band