期刊文献+

医疗设备开关电源典型故障分析与维修 被引量:8

Analysis and Maintenance of Typical Faults of Switching Power Supply for Medical Equipment
下载PDF
导出
摘要 安全稳定的开关电源是保证设备正常运行的前提条件。本文总结HVS-2型口腔专用负压机、二氧化碳培养箱Thermo8000DH、迈瑞T5监护仪三个典型设备电源故障实例,深入解析电源开关的组成及工作原理,根据分析结果制定相应的解决方案。通过对开关电源故障的分析与检修,故障得以处理,为医疗设备开关电源的维修提供了实例参考。同时,只有维修人员充分掌握了医疗设备开关电源的原理与检修方法,才可能为医疗设备的安全稳定运行提供保障。 Safe and stable switching power supply is the prerequisite to ensure the normal operation of the equipment.Three typical equipment power failure examples of HVS-2 oral negative pressure machine,carbon dioxide incubator Thermo8000DH and Mindray T5 monitor were summarized in this article,the composition and working principle of the power switch was thoroughly analyzed,and the corresponding solution was formulated based on the analysis results.We analyzed and repaired the fault of switching power supply,so that the fault could be handled,which provided a reference for the maintenance of switching power supply of medical equipment.At the same time,only when the maintenance personnel have fully mastered the principle and maintenance methods of the switching power supply of medical equipment,can they provide guarantee for the safe and stable operation of medical equipment.
作者 夏信堂 刘勇 李奥林 单长征 卞兆峰 李云雷 田野 蔡杨 XIA Xintang;LIU Yong;LI Aolin;SHAN Changzheng;BIAN Zhaofeng;LI Yunlei;TIAN Ye;CAI Yang(Department of Equipment,Qilu Children’s Hospital of Shandong University,Jinan Shandong 250000,China)
出处 《中国医疗设备》 2020年第3期167-170,共4页 China Medical Devices
关键词 开关电源 口腔专用负压机 二氧化碳培养箱 监护仪 故障分析 switching power supply oral negative pressure machine carbon dioxide incubator monitor fault analysis
  • 相关文献

参考文献17

二级参考文献68

共引文献43

同被引文献88

引证文献8

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部