期刊文献+

TASwako i30全自动免疫仪检测仪性能评价

Performance Evaluation of μTASwako i30 Automatic Immunodetector
下载PDF
导出
摘要 目的 通过评估TASwako i30全动免疫检测仪的性能,以确保实验数据的准确.方法 分别从精密度,线性范围,准确性,室间比对方面对TASwako i30进行全面评价.结果 批内精密度<5%、批间精密度<10%、均在允许范围内.线性:参照CLSI的EP-6A2文件进行线性评估良好,准确性良好,与其他医院对比结果符合标准.结论 μTASwako i30全自动免疫检测仪性各项性能参数结果符合标准,适合用于临床. Objective By assessing TASwako i30 dynamic immune detector performance,and to ensure the accuracy of experimental data.Methods Respectively from the precision,linear range,accuracy,and comparison between external of TASwako i30 comprehensive evaluation.Results Within-run precision<5%,between-run precision<10%,all within the permitted range.Linear evaluation is good by referring to the EP-6 A2 file of CLSI,good accuracy.Compared with other hospitals,the results met the standard.Conclusion μTASwako i30 automatic immunity analyzer the results of various performance parameters accord with the standards and are suitable for clinical use.
作者 徐建 许亮亮 喻海忠 Jian XU;Liangliang XU;Haizhong YU(Nantong Hospital of Traditional Chinese Medicine,Nantong Jiangsu 226001,China)
机构地区 南通市中医院
出处 《国际感染病学(电子版)》 CAS 2019年第3期7-9,共3页 Infection International(Electronic Edition)
关键词 全自动免疫检测仪 性能评价 精密度 Automatic immunity analyzer Performance evaluation Precision
  • 相关文献

参考文献3

二级参考文献27

  • 1蔡明翠.血站检验实验室的质量控制[J].职业卫生与病伤,2006,21(3):214-215. 被引量:5
  • 2杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127. 被引量:173
  • 3Xu Fen, Luo Jie, Li Yi.college testing laboratory management andsecurity[J]. HEALTH DIGEST, 2011 (4); 89-92. 被引量:1
  • 4Liu xue yong. Quality management of clinical testing laboratories[J].GUIDE OF CHINA MEDICINE, 2009, 7 (3): 156-158. 被引量:1
  • 5Cao Jun .Quality management of blood bank blood testinglaboratories[J]. CHINA MEDICAL HERALD, 2008,5 ( 30 ) : 99-101. 被引量:1
  • 6朴美花.浅谈临床检验标本的质量控制与管理对策[J].吉林医学,2008,29(1):75-75. 被引量:10
  • 7中华人民共和国卫生部.WS/T 347-2011血细胞分析的校准指南[S].北京:中国标准出版社,2008. 被引量:1
  • 8刘志浩,黄文源.实用临床血液学图谱[M].北京:科学出版社,1996. 被引量:1
  • 9Inagaki Y, Tang W, Makuuchi M, et al. Clinical and molecular insights into the hepatocellular carcinoma tumour marker des-gam- ma-carboxypmthrombin[J]. Liver Int, 2011, 31 (1): 22-35. 被引量:1
  • 10Lok AS, Sterling RK, Everhart JE, et al. Des-gamma-carboxy prothmmbin and alpha-fetoprotein as biomarkers for the early de- tection of hepatocellular carcinoma[J]. Gastroenterology, 2010, 138(2) :493-502. 被引量:1

共引文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部