摘要
目的 通过评估TASwako i30全动免疫检测仪的性能,以确保实验数据的准确.方法 分别从精密度,线性范围,准确性,室间比对方面对TASwako i30进行全面评价.结果 批内精密度<5%、批间精密度<10%、均在允许范围内.线性:参照CLSI的EP-6A2文件进行线性评估良好,准确性良好,与其他医院对比结果符合标准.结论 μTASwako i30全自动免疫检测仪性各项性能参数结果符合标准,适合用于临床.
Objective By assessing TASwako i30 dynamic immune detector performance,and to ensure the accuracy of experimental data.Methods Respectively from the precision,linear range,accuracy,and comparison between external of TASwako i30 comprehensive evaluation.Results Within-run precision<5%,between-run precision<10%,all within the permitted range.Linear evaluation is good by referring to the EP-6 A2 file of CLSI,good accuracy.Compared with other hospitals,the results met the standard.Conclusion μTASwako i30 automatic immunity analyzer the results of various performance parameters accord with the standards and are suitable for clinical use.
作者
徐建
许亮亮
喻海忠
Jian XU;Liangliang XU;Haizhong YU(Nantong Hospital of Traditional Chinese Medicine,Nantong Jiangsu 226001,China)
出处
《国际感染病学(电子版)》
CAS
2019年第3期7-9,共3页
Infection International(Electronic Edition)
关键词
全自动免疫检测仪
性能评价
精密度
Automatic immunity analyzer
Performance evaluation
Precision