期刊文献+

平板式脱硝催化剂化学成分测试中铬元素干扰消除的研究

Study on Elimination of Chromium Interference in Chemical Composition Test of Plate Denitrification Catalyst
下载PDF
导出
摘要 采用X射线荧光光谱法测试平板式脱硝催化剂化学成份时,因催化剂的有效成分与基材剥离时,会把基材中的干扰元素引入到有效成分中,特别是基材中的铬与钒具有相似的波长,影响测试。文章采用磁铁吸附法、波长调整法两种手段,消除了铬对钒测试的影响,测试数据准确、可靠。 When X-ray fluorescence spectrometry is used to measure the chemical composition of flat-plate denitrification catalyst, the effective component of catalyst and the base material will be introduced into the effective component, especially when the chromium and vanadium in the base material have similar wavelength, which will affect the test. In this paper, the influence of chromium on vanadium measurement is eliminated by magnet adsorption method and wavelength adjustment method. The test data are accurate and reliable.
作者 闫力 YAN Li(Datang Nanjing Environmental Protection Technology Co.,Ltd.,Nanjing 211111,China)
出处 《化工管理》 2019年第34期132-134,共3页 Chemical Engineering Management
关键词 XRF 铬元素 消除干扰 XRF chromium eliminate interference