基于ExacTrac X射线图像引导系统的无创颅脑SRS精确性研究 被引量:2

Accuracy of intracranial non-invasive stereotactic radiotherapy surgery based on ExacTrac X-ray image guidance system
下载PDF
导出
摘要 目的:探讨无创颅脑立体定向放射外科(stereotactic radiotherapy surgery,SRS)中ExacTrac X射线图像引导系统纠正摆位误差的精确性。方法:选择30例颅脑肿瘤患者,在SRS治疗前用ExacTrac X射线图像引导系统验证,利用六维床校正Lat(左右)、Long(头脚)、Vert(前后)3个平移误差和Lat Ang、Long Ang、Vert Ang 3个旋转误差,并在校正后的位置上再拍一组X射线片与数字重建放射影像(digitally reconstructured radiograph,DRR)匹配,记录6个方向的误差值,分析校正前后6个方向的均值和标准差。匹配完成后进行锥形束CT(cone beam CT,CBCT)扫描,医生在线匹配CBCT图像与定位CT图像,得到Lat、Long、Vert 3个平移误差,将其与经ExacTrac X射线图像引导系统校正后的3个平移误差进行比较,分析2个系统的设置差异。结果:校正前Lat、Long、Vert 3个平移误差平均值/标准差分别为-0.42/1.15 mm、0.72/1.29 mm、-0.49/1.13 mm,Lat Ang、Long Ang、Vert Ang 3个旋转误差均值/标准差分别为-0.34/0.83°、-0.11/1.01°、0.30/1.33°。校正后6个方向的误差明显缩小,3个平移方向的误差均≤0.5 mm,3个旋转方向的误差均<0.5°。3个平移方向和Lat Ang旋转方向误差差异具有统计学意义(P<0.05),在Long Ang、Vert Ang 2个旋转方向误差差异不具有统计学意义(P>0.05)。CBCT与ExacTracX射线图像引导系统2个系统3个平移方向误差的差值均≤2 mm。结论:应用ExacTrac X射线图像引导系统配合六维床可明显减小误差值,说明在颅脑SRS治疗中应用ExacTrac X射线图像引导系统是很有必要的。 Objective To investigate the accuracy of ExacTrac X-ray image guidance system and compare the setup errors with ExacTrac X-ray image guidance system and cone-beam computed tomography(CBCT) for intracranial non-invasive stereotactic radio-therapy surgery(SRS). Methods Totally 30 craniocerebral tumor patients underwent pre-SRS verification by ExacTrac X-ray image guidance system, and the 6 D bed was used to correct the translation and rotation errors at the directions of Lat(left-right), Long(head-feet) and Vert(forward-backward) as well as Lat Ang, Long Ang and Vert Ang. Then another group of X-ray images were taken at the corrected position and matched with DRR images, and the values were recorded at the six directions. The mean values and standard deviations were analyzed before and after correction. After the matching was completed cone beam CT(CBCT) scan was performed. The doctor matched the CBCT image and the CT image online, and obtained the translation errors at Lat, Long and Vert. The translation errors obtained were compared with those undergoing correction by Exactrac X-ray image guidance system to analyze the setup difference between the two systems.Results Before correction the mean errors and standard deviations were respectively-0.42/1.15 mm, 0.72/1.29 mm and-0.49/1.13 mm at Lat, Long and Vert directions, and-0.34/0.83°,-0.11/1.01° and 0.30/1.33° respectively at Lat Ang, Long Ang and Vert Ang directions. After correction the errors decreased significantly at the six directions, which were not more than 0.5 mm at the translation directions and less than 0.5° at the rotation directions. There were statistical differences for the errors at the three translation directions and Lat Ang directions(P<0.05), while no for those at Long Ang and Vert Ang directions(P>0.05). The two systems had all the differences between the errors at the three translation directions not more than 2 mm.Conclusion ExacTrac X-ray image guidance system combined with 6 D bed decreases the error significantly, and thus gains nec
作者 谭薇薇 罗灿 周军 TAN Wei-wei;LUO Can;ZHOU Jun(The Second Affiliated Hospital of Army Medical University,Chongqing 400037,China)
出处 《医疗卫生装备》 CAS 2019年第7期44-48,共5页 Chinese Medical Equipment Journal
关键词 ExacTrac X射线图像引导系统 立体定向放射治疗 六维床 摆位误差 CBCT ExacTrac X-ray image guidance system stereotactic radiotherapy 6D bed setup error CBCT
  • 相关文献

参考文献2

二级参考文献41

  • 1杨波,赵涤非,张福正.改良放疗体膜固定装置的摆位[J].江苏大学学报(医学版),2005,15(4):335-336. 被引量:2
  • 2邵明海,胡炜,余文杰,于长辉,郑永敏.胸腹体架热塑体膜固定摆位误差分析[J].浙江临床医学,2006,8(4):433-434. 被引量:8
  • 3吴云来,赵家成,张雷,李文海,江浩.头颈部肿瘤外照射面罩体位固定技术应用价值的研究[J].实用全科医学,2006,4(5):537-538. 被引量:10
  • 4Stine Korreman,科恩罗序,海伦捷.等.欧洲放射治疗及肿瘤协会和欧洲放疗研究所关于基于三维CT的影像引导系统报告:一种实用的技术回顾和指南[J].放疗和肿瘤,2010,94(2):129-144. 被引量:1
  • 5W.H.格林.S.Chelikani, K. Purushothaman, 等.约束非刚性配准在图像引导的自适应放疗中的使用[J].医学图像分析,2009,13(50):809-817. 被引量:1
  • 6O.FLUCK,C.维特尔,W.Wein,等.图形硬件上医学图像配准的调查[J].生物医学计算机方法和程序,2010. 被引量:1
  • 7丹尼尔 Rui-jters,哈尔 Romeny,保罗 Suetens.应用于内手术的GPU加速三维图像弹性配准[J].生物医学计算机方法和程序,2010. 被引量:1
  • 8罗布 H.爱尔兰.卡伦 E Dyker,大卫 CBarber,等.头颈部癌症放射治疗计划中PET/CT的非刚性图像配准[J].国际放射肿瘤生物物理杂志,2007,68(3):952-957. 被引量:1
  • 9彼得 J.Slomka,理查德 P 鲍姆.多模态图像配准软件:最先进的技术[J].欧州核医学分子影像杂志,2009,36(增刊):S44-S55. 被引量:1
  • 10He 王,苏尼尔 克里希南,王晓春,等.使用变形图像配准改善肝癌患者四维CT图像的软组织对比度[J].国际放射肿瘤生物物理杂志,2008,72(1):201-209. 被引量:1

共引文献21

同被引文献22

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部