移液器容量自动校准系统研究 被引量:6

Research on Automatic Calibrating System for Pipette Capacity
下载PDF
导出
摘要 针对移液器人工校准中存在的按压不稳、温度偏差、数据无法批量处理等缺陷,提出了一种基于模拟人工按压的移液器容量自动校准方法。通过测量移液器按钮2个停止点的压力值,获取近似按压的距离;分析数据并与人工操作进行比对,得出模拟手动按压的最佳按压距离。开发了一套同时校准4支移液器容量的自动化装置,采用机械手和Lab VIEW软件智能控制取液、排液、称量、数据处理等工作,实现了自动校准。研究结果表明,与人工操作相比,重复性减小了0. 0%~0. 9%,相对误差减小了0. 1%~0. 7%。 Aim at the shortcomings of manual calibration such as instability of the press, temperature deviation and data that can not been processed in batches, an automatic calibration method for pipette capacity based on simulated manual press was proposed. By measuring the pressure at the two stop points of the button of the pipette, the approximate pressing distance can be obtained;comparing and analyzing data with manual operation, the optimum press distance of simulated manual press was obtained. An automatic device for simultaneous calibration of 4 pipettes was developed. Those works were intelligently controlled including extraction, discharge, weighing, and data processing adopting manipulator and LabVIEW software, fully-automatic calibrating were realized. Research results show that comparing to manual operation, the repeatability reduces 0.0%~0.9%, and the relative error decreases by 0.1%~0.7%.
作者 赵玉晓 谢玄达 孙斌 张竟月 文英杰 ZHAO Yu-xiao;XIE Xuan-da;SUN Bin;ZHANG Jing-yue;WEN Ying-jie(College of Modern Science and Technology, China Jiliang University, Hangzhou, Zhejiang 310018, China;National Institute of Metrology, Beijing 100029, China;Hangzhou Institute of Calibration andTesting for Quality and Technical Supervision, Hangzhou, Zhejiang 310019, China)
出处 《计量学报》 CSCD 北大核心 2019年第3期392-396,共5页 Acta Metrologica Sinica
基金 国家自然科学基金(51475440) 国家质量监督检验检疫总局科技计划项目(2017QK123)
关键词 计量学 移液器 自动校准 小容量计量 不确定度 metrology pipette automatic calibrating small capacity metrology uncertainty
  • 相关文献

参考文献3

二级参考文献12

  • 1JJF1059-1999.测量不确定度评定与表示[S].[S].,.. 被引量:779
  • 2JJF1059-1999.测量不确定度评定与表示[S].[S].,1999.. 被引量:183
  • 3Tanaka, M.,et. Al. Recommended table for the density of waterbetween 0 X. and 40? based on recent experime ntalreports. Metrologia, 2001,38:301 -309. 被引量:1
  • 4ISO4787 :2010 Laboratory glassware - Volumetric glassware - Meth-ods for use and testing of capacity[ S]. 被引量:1
  • 5ISO8655 -1/2/6:2002 Piston - operated volumetric apparatus[ S]. 被引量:1
  • 6S Lorefice. Traceability and uncertainty analysis in volume measure-ments [J ]. Measurement, 2009 ,42 :1510-1515. 被引量:1
  • 7Tanaka M, et al. Recommended table for the density of water between 0C and 40C based on recent experimental reports [ J ]. Metrologia ,2001,38:301 - 309. 被引量:1
  • 8The International Organlzation for Standardization ISO4787 : 2010, Laboratory glassware-Volumetric glassware-Methods for use and testing of capacity [ S ]. 2010. 被引量:1
  • 9The International Organization for Standardization ISO8655-1/2/6 : 2002, Piston-operated volumetric apparatus [ S ]. 2002. 被引量:1
  • 10Lorefice S. Traceability and uncertainty analysis in volume measurements [ J ]. Measurement ,2009,42 : 1510 - 1515. 被引量:1

共引文献5

同被引文献24

引证文献6

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部