期刊文献+

便携式免散瞳带信标眼底相机光学系统 被引量:2

A portable and mydriasis-free fundus optical system with a beacon
原文传递
导出
摘要 设计了一款便携式39°视场免散瞳眼底相机光学系统,可以对-10~+15m^(-1)屈光度变化范围内的人眼视网膜清晰成像。采用环形光源照明,使眼底视网膜照明均匀度达到95%,也很好地消除了角膜反射。加入的信标系统使得便携式眼底相机光学系统在拍摄时有助于保持人眼与成像系统的同轴度,进一步提高拍摄视网膜图像的清晰度。通过对成像系统的优化设计及仿真分析,得到均方根直径小于6.4μm、场曲小于60μm、像散小于50μm、畸变小于3.6%,在67lp/mm处各视场传递函数值大于0.5和成像系统加工良率高的眼底相机光学系统方案。方案可以应用于眼科检查,改善全国偏远地区眼科医疗器械缺少的情况。 A portable and mydriasisfree fundus optical system with 39° field of view(FOV) is designed. Legible retina images of eyes in which vergence ranges from-10^+15 m-1 can be received by this system. An annular light source is adopted to effectively eliminate the stray light reflected by cornea and the retinal illumination uniformity can be up to 95%. The retinal image quality can be improved by the beacon system, which contributes to maintain the coaxiality of the human eye with the imaging system while the fundus optical system is working. Through optimization design and simulation analysis of the imaging system, a fundus optcial system with great fabrication feasibility is designed. Results show that the modulation transfer function(MTF) values of all fields are higher than 0.5 at 67 lp/mm with root-mean-square(RMS) less than 6.4μm, the field curve is less than 60μm, the astigmatism is less than 50μm and the distortion is less than 3.6%. The program can be applied to improve the lack of ophthalmic medical devices in remote areas of the country.
作者 江剑宇 杨波 万新军 张薇 韦晓孝 张婧京 黄鹏程 JIANG Jianyu;YANG Bo;WAN Xinjun;ZHANG Wei;WEI Xiaoxiao;ZHANG Jingjing;HUANG Pengcheng(School of Optical-electrical and Computer Engineering, Shanghai Key Lab of Modem Optical System Research Center of Optical Instruments and System Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China)
出处 《光学技术》 CAS CSCD 北大核心 2019年第2期240-244,共5页 Optical Technique
基金 国家重点研发计划(2016YFB0402000)
关键词 眼底相机光学系统 照明系统 成像系统 信标系统 fundus optical system illumination system imaging system beacon system
  • 相关文献

参考文献6

二级参考文献64

  • 1凌宁,张雨东,饶学军,李新阳,王成,胡弈云,姜文汉.用于活体人眼视网膜观察的自适应光学成像系统[J].光学学报,2004,24(9):1153-1158. 被引量:45
  • 2萧泽新.眼底电视光学系统的设计[J].光学技术,2005,31(2):170-172. 被引量:7
  • 3郭欢庆,王肇圻,赵秋铃,全薇,王雁.基于主观式测量人眼波前像差的眼模型研究[J].光子学报,2005,34(11):1666-1669. 被引量:15
  • 4Hubbard LD,Brothers RJ,King WN,et al.Methods for evaluation of retinal microvascular abnormalities associated with hypertension/sclerosis in the Atherosclerosis Risk in Communities (ARIC)Study.Ophthalmology,1999,106:2269-2280. 被引量:1
  • 5Wong TY,Klein R,Sharrett AR,et al.Retinal arteriolar narrowing and risk of diabetes mellitus in middle-aged persons.JAMA,2002,287:2528-2533. 被引量:1
  • 6Wong TY,Shankar A,Klein R,et al.Retinal arteriolar narrowing,hypertension,and subsequent risk of diabetes mellitus.Arch Intern Med,2005,165:1060-1065. 被引量:1
  • 7Ikram MK,Janssen JA,Roos AM,et al.Retinal vessel diameters and risk of impaired fasting glucose or diabetes:the Rotterdam Study.Diabetes,2006,55:506-510. 被引量:1
  • 8Nguyen TT,Wang JJ,Wong TY.Retinal vascular changes in prediabetes and prehypertension:New findings and their research and clinical implications.Diabetes Care,2007,30:2708-2715. 被引量:1
  • 9Klein R,Klein BE,Moss SE,et al.Retinal vascular abnormalities in persons with type Ⅰ diabetes:The Wisconsin Epidemiologic Study of Diabetic Retinopathy:ⅩⅧ.Ophthalmology,2003,110:2118-2125. 被引量:1
  • 10Klein R,Klein BE,Moss SE,et al.The relation of retinal vessel caliber to the incidence and progression of diabetic retinopathy:ⅩⅨ.The Wisconsin Epidemiologic Study of Diabetic Retinopathy.Arch Ophthalmol,2004,122:76-83. 被引量:1

共引文献29