期刊文献+

测量350~1100nm光源光谱总辐射通量的标准装置 被引量:2

The Standard Measurement Facility for Total Spectral Radiant Flux Measurement of 350~1100nm Light Source
下载PDF
导出
摘要 为了满足快速发展的LED行业对光谱总辐射通量的计量需求,研制了一套用于测量光谱总辐射通量的积分球光谱辐射计装置,实现光谱总辐射通量量值的准确传递;介绍了该装置的基本原理、操作规范和不确定度评估。经实例评估,该装置在350~1 100 nm的扩展测量不确定度为U_(rel)=3.1%~1.3%(k=2)。对一组光谱总辐射通量标准灯与变角辐射计进行比对测量,实验结果表明在350~1 100 nm波长上该标准灯组内一致性偏差为0.2%~0.8%,能满足量值传递的需求。 To meet the increasing demand for total spectral radiant flux (TSRF) measurement in the LED industry, we developed a facility of an integrating sphere with a spectroradiometer, and developed relative standard operational procedure. The facility can quickly transfer the TSRF quantity from the standard lamp to the testing lamp. The basic concept, standard operational procedure and measurement uncertainty analysis are shown. The TSRF expanded uncertainty at wavelength range 350~1 100 nm is U rel =3.1%~1.3%( k =2 ). A group of TSRF standard lamp, calibrated by the goniospectroradiometer, are tested using this facility. The experimental results showed the consistency deviations of the group of standard lamp were small than 0.8% at the wavelength range of 350~1 100 nm.
作者 赵伟强 刘慧 闫劲云 苏颖 姜琳 ZHAO Weiqiang;LIU Hui;YAN Jingyun;SU Ying;JIANG Ling(National Institute of Metrology,Beijing 100029,China)
出处 《照明工程学报》 2019年第2期4-7,共4页 China Illuminating Engineering Journal
基金 国家自然科学基金(批准号:61505191) 中国计量科学研究院项目(批准号:ANL1605)
关键词 光谱总辐射通量 积分球光谱辐射计 光谱辐射计 标准灯 计量 total spectral radiant flux spherical spectroradiometer spectroradiometer standard lamp metrology
  • 相关文献

参考文献5

二级参考文献17

  • 1吕正.LED光通量测试中的若干问题[J].中国照明电器,2005(12):29-30. 被引量:11
  • 2Ohno Y, Zong Y. NIST Facility for Total Spectral Radiant Flux Calibration [ C ]//Symposium of Metrologia. CENAM, Mexico : BIPM,2004,1-6. 被引量:1
  • 3Zong Y, Ohno Y. Realization of the NIST Total Spectral Radiant Flux Scale[ C] // Proceedings of 26th Session of the CIE. Beijing China: CIE, 2007, 179 - 182. 被引量:1
  • 4Ohno Y. Realization of NIST 1995 Luminous Flux Scale using Integrating Sphere Method[ J ]. IES, 1996,25 (1) : 13 -22. 被引量:1
  • 5Shaw M, Goodman T. Array-based goniospectroradiometer for measurement of spectral radiant intensity and spectral total flux of light sources [ J ]. Applied Optics, 2008, 47 ( 14), 2637 - 2647. 被引量:1
  • 6Zong Y, Brown S W, Johnson B C, et al. Simple spectral stray light correction method for array spectroradiometers [ J 1. Appl Opt, 2006, 45 ( 6 ) : 1111 - 1119. 被引量:1
  • 7潘建根,李倩,伍德辉,等.GB/T 24824-2009普通照明用LED模块测试方法[S].北京:中国标准出版社,2010. 被引量:1
  • 8Tomas D,Paolo T,Lukas V,et al.Comparison of constanst and temperature dependent blood perfusion in temperature prediction for superficial hyperthermia[J].Radio Engineering,2010,19(2):281-289. 被引量:1
  • 9李振国,王岑娥,徐晓光.一种用积分球实现总光通量标的新方法[J].现代计量测试,1998,6(2):59-61. 被引量:2
  • 10林晓珑,张铁强,郭山河.汽车车灯光通量测试系统的研制[J].光学精密工程,1998,6(4):132-134. 被引量:3

共引文献39

同被引文献20

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部