口腔综合治疗台治疗用水污染危险因素调查 被引量:5

Risk Factor of Contamination in Dental Unit Waterlines
下载PDF
导出
摘要 目的了解上海市徐汇区口腔综合治疗台治疗用水污染状况,并分析污染的危险因素。方法采集上海市徐汇区9家医疗机构34台综合治疗台不同时段三种管道的治疗用水,进行细菌菌落总数计数,同时收集相关资料进行危险因素分析。结果上午开诊前、第一个病人诊疗后、上午诊疗结束后治疗用水合格率分别为57.6%,70.7%,87.0%。综合治疗台当日的使用时间短是影响治疗用水菌落总数的主要因素。治疗用水合格率的危险因素分析中,回吸强度大、采用储水罐供水方式、水处理设备管理水平差及当日管道冲洗时间短都是独立的危险因素。结论口腔综合治疗台治疗用水污染情况严重,存在引起医院感染的隐患。影响污染状况的因素较多,需要有综合措施对综合治疗台水路污染状况进行控制。 Objective To investigate the contamination of dental unit waterlines and analyze the associated risk.METHODS Water samples were collected from 34 dental unit of 9 hospital in Shanghai,and the TBC counted.Meanwhile,the associated factors of contamination were collected.RESULTS The qualification rate of water samples collected in different time were 57.6%,70.7%,87.0%.The short time of using wateline system was the main factor influenced to the total bacterial counts of water samples.In the risk factors analysis for contamination of DUWS,more sucking power,tank-fed water supply system,poor management of waterline,short flushing time were independent risk factors.CONLUSION Most of the investigated dental unit have high level contamination,which would likely to acquire nosocomial infection.The complex of risk factors influence contamination of DUWS,the comprehensive treatments should be accomplished to control the contamination.
作者 张玲 张越文 闵红芳 孙冬梅 钱子煜 Zhang Lin;Zhang Yue-wen;Min Hong-fang;Sun Dong-mei;Qian Zi-yu(Shanghai Xuhui Dental Disease Prevention and Control Institute,Shanghai 200237,China;Disinfection Department,Shanghai Xuhui Disease Prevention and Control Center,Shanghai 200237,China;Microbiological Laboratory,Shanghai Xuhui Disease Prevention and Control Center,Shanghai 200237,China)
出处 《全科口腔医学电子杂志》 2019年第2期1-4,共4页 Electronic Journal of General Stomatology
基金 2017年徐汇区医学尖峰学科(SHXH201706)
关键词 综合治疗台水路 污染 医院感染 危险因素 Dental unit waterlines Contamination Hospital acquired infection Risk factors
  • 相关文献

参考文献6

二级参考文献50

  • 1张莉萍,张勤英,张蕾薇.口腔科控制涡轮手机媒介交叉感染的难点和对策[J].护理学报,2006,13(8):38-39. 被引量:5
  • 2Montebugnoli L,Sambri V,Cavrini F,et al.Detection of DNA from periodontal pathogenic bacteria in biofilm obtained from waterlines in dental units.New Microbiol,2004,27 (4):391-397. 被引量:1
  • 3Szymanska J.Electron microscopic examination of dental unit waterlines biofilm.Ann Agric Environ Med,2005,12(2):295-298. 被引量:1
  • 4Mavridou A,Kamma J,Mandilara G,et al.Microbial risk assessment of dental unit water systems in general dental practice in Greece.Water Sci Technol,2006,54(3):269-273. 被引量:1
  • 5ADA Council on Scientific Affairs and ADA Council on Dental Practice.Infection control recommendations for the dental office and the dental laboratory.J Am Dent Assoc,1996,127(5):672-680. 被引量:1
  • 6Souza-Gugelmin MC,Lima CD,Lima SN,et al.Microbial contamination in dental unit waterlines.Braz Dent J,2003,14 (1):55-57. 被引量:1
  • 7Mills SE.Waterbome pathogens and dental waterlines.Dent Clin North Am,2003,47(3):545-557. 被引量:1
  • 8Kettering JD,Mufloz-Viveros CA,Stephens JA,et al.Reducing bacterial counts in dental unit waterlines:distilled water vs.antimicrobial agents.J Calif Dent Assoc,2002,30 (10):735-741. 被引量:1
  • 9Mills SE,Karpay RI.Dental waterlines and biofilm-searching for solutions.Compend Contin Educ Dent,2002,23 (3):237-240,242,244. 被引量:1
  • 10Prevost AP,Robert M,Charland R,et al.Doctor,would you drink water from your dental unit? N Y State Dent J,1995,61 (10):22-28. 被引量:1

共引文献83

同被引文献36

引证文献5

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部