期刊文献+

HM-200型移动式高频X线机灯丝回路故障检修一例 被引量:1

One Case of Fault Diagnosis of Filament Circuit for HM-200 Mobile High-Frequency X-ray Machine
下载PDF
导出
摘要 本文通过分析检修HM-200型移动式高频X线机一例灯丝回路故障,为X线机维修提供借鉴经验。对于出现的灯丝回路故障,根据其工作原理逐步分析,使用示波器捕捉其波形,初步确定故障范围,分析电路图查找故障点,最终精确到故障元件,更换故障元件排除故障。该维修方法可行,且维修成本极低。 This article aimed to provide experience for X-ray machine maintenance based on analyzing and solving a case of HM-200 mobile high-frequency X-ray machine filament circuit fault.For the filament loop fault,we firstly analyzed step by step according to the working principle,then the fault range was determined by using oscilloscope to capture the waveform.By analyzing the circuit diagram to find the fault point,the fault element was finally located accurately,and the fault element was replaced to eliminate the fault.This method of maintenance is feasible and the cost is very low.
作者 朱明明 秦志强 赵先东 宋丰言 ZHU Mingming;QIN Zhiqiang;ZHAO Xiandong;SONG Fengyan(Department of Radiology,Xining Joiot Logistics Support Center Drug Instrument Supervision and Inspection Station,Chengdu Sichuan 610017,China)
出处 《中国医疗设备》 2019年第1期163-165,共3页 China Medical Devices
关键词 HM-200型移动式高频X线机 灯丝回路 故障维修 曝光故障 HM-200 mobile high-frequency X-ray machine filament loop malfunction repair exposure trouble