能谱CT单能量成像结合低浓度对比剂在门静脉成像的应用价值 被引量:11

Application Value of Energy Spectrum CT Single Energy Imaging Combined with Low Concentration Contrast Agent in Portal Vein Imaging
下载PDF
导出
摘要 目的探讨能谱CT单能量成像结合低浓度对比剂在门静脉成像的应用价值。方法 70例欲行肝门脉CT增强患者随机分为A、B两组。A组用CT能谱成像(Gemstone Spectral Imaging,GSI),270 mg I/m L碘克沙醇;B组用常规CT增强扫描,350 mg I/m L碘佛醇。计算两组门静脉图像噪声、对比噪声比(Contrast-to-Noise Ratio,CNR)、信噪比(Signal-to-Noise Ratio,SNR)、图像主观评分及有效辐射剂量(ED),并进行相应统计学分析。结果 A组门静脉最佳CNR图像单能量水平在50 Ke V左右,最佳单能图结合50%自适应统计迭代重建(Adaptive Statistical Iterative Recon,ASi R)图像的SNR(7.46±1.00)、CNR(5.31±1.23)均高于B组SNR(5.03±1.00)、CNR(3.18±1.58);噪声(16.58±7.16)和平均ED(3.27±1.57)m Sv低于B组噪声(23.19±11.71)及平均ED(4.06±0.85)m Sv,差异具有统计学意义。A组图像主观评分略高于B组,差异不具有统计学意义。结论能谱CT最佳单能量成像结合低浓度对比剂,采用ASi R技术,得到的门静脉图像质量与常规CT增强相当,辐射剂量降低,值得在临床上推广。 Objective This paper aim to explore the application value of energy spectrum CT single energy imaging combined with low concentration contrast agent in portal vein imaging.Method70patients were randomly divided into A,B two groups to receive liver CT enhanced scan.A group using the gem CT energy spectrum imaging(GSI),270mgI/mL iodixanol;B group using conventional CT enhanced scanning,350mgI/mL ioversol.Calculation the portal image noise,contrast to noise ratio(CNR),signal to noise ratio(SNR),image quality subjective evaluation and the effective radiation dose(ED)of the two groups,and the corresponding statistical analysis.Results The SNR(7.46±1.00)and CNR(5.31±1.23)of images in the A group were higher than those of the B group[(5.03±1.000)and(3.18±1.58)].The noise(16.58±7.16)of image and average ED(3.27±1.57)mSv were lower than those of the B group[(23.19±11.71)and(4.06±0.85)],the difference was statistically significant.The subjective scores of images in the A group were slightly higher than those in the B group,but there was no statistical difference between the two groups.Conclusion The best single energy spectrum CT imaging combined with low concentration contrast,using adaptive statistical iterative reconstruction(ASiR)technology,it can get the same image quality as the conventional enhanced CT,reduce radiation dose,is worth to be popularized in clinic.
作者 周泽旺 张昌政 郑瑛琪 李丹丹 ZHOU Ze-wang;ZHANG Chang-zheng;ZHENG Ying-qi;LI Dan-dan(Department of Radiology, the First Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou Guangdong 510405, China)
出处 《中国医疗设备》 2017年第5期87-90,共4页 China Medical Devices
关键词 能谱单能量 门静脉 低浓度对比剂 X线计算机 辐射剂量 energy spectrum single energy portal vein low concentration contrast agent X-ray computer radiation dosage
  • 相关文献

参考文献9

二级参考文献55

  • 1王家臣,刘学静,田军,巩若箴.多层螺旋CT肝门静脉成像对门脉高压分流侧支血管的研究[J].医学影像学杂志,2007,17(9):991-994. 被引量:11
  • 2Matsumoto K, Jinzaki M, Tanami Y, et al. Virtual monochromatic spectral imaging witil fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology ,2011,259:257-262. 被引量:1
  • 3Zhao LQ, He W, Li JY, et al. Improving image quality in portal venography with spectral CT imaging. Eur J Radiol, 2012, 81 : 1677-1681. 被引量:1
  • 4Lv P, Lin XZ, Li J, et al. Differention of small hepatic hemangioma from small hepatocellular carcinoma: recently introduced spectral CT method. Radiology ,2011,259:720-729. 被引量:1
  • 5Zhu K, et al. Gastric varices in patients with portal hyperten- sion: evaluation with multideteetor row CT [J]. J Clin Gas- troenterol, 2010, 44: 108-115. 被引量:1
  • 6Agarwal A, Jain M. Multidetector CT portal venography in evaluation of portosystemie collateral vessels [J]. J Med Ima-ging Radiat OncoI, 2008, 52: 4-9. 被引量:1
  • 7Li J, et al. An optimal contrast dose indicator for the deter- mination of hepatic enhancement in abdominal multidetector computed tomography: comparison of patient attenuation in dicator with total body weight and body mass index [J]. J Comput Assist Tomogr, 2010, 34: 874-878. 被引量:1
  • 8Gleeson TG, Bulugahapitiya S. Contrast-induced nephropa- thy[J]. AJR, 2004, 183: 1673-1689. 被引量:1
  • 9Brooks RA. A quantitative theory of the hounsfield unit and its application to dual energy scanning[J]. J Comput Assist Tom0gr, 1977, 1: 487-493. 被引量:1
  • 10刘爱连,沈云.能谱CT临床应用图谱[M].北京:人民军医出版社,2012:92-93. 被引量:11

共引文献159

同被引文献100

引证文献11

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部