期刊文献+

三浮陀螺仪沿输出轴误差系数D_o的产生机理分析 被引量:1

Mechanism Analysis of SDFG's Coefficient Attributable to Acceleration along the Output Axis
原文传递
导出
摘要 目前,三浮陀螺仪误差模型包含沿输出轴的一次项误差系数D_0,该误差项尚无明显的物理意义,但|D_0|的标定结果在0.0002(°)/(h·g)~0.015(°)/(h·g)。针对三浮陀螺仪的结构特点和测试方法分析了D_0的产生机理,通过调整安装误差角的试验说明安装误差角带来的天向地速和一次项分量会影响D_0,因此系统在选用陀螺误差模型时,需关注由测试方法带来的误差。同时,陀螺在设计、生产和调试过程中对安装误差角的影响应予以重视。最后用统计检验的方法说明D_0对误差模型的线性影响不显著,为系统选取陀螺误差模型提供依据。 Nowadays, SDFG's application model equation is including coefficient D_0 which is sensitive times to OA acceleration. There is no exactly explanation for the coefficient which is in the range of 0.0002(°)/(h · g)~0.015(°)/(h · g) according to the testing results. The analysis of this coefficient D_0 based on the SDFG's mechanism structure and testing method are discussed in this paper. By adjusting the misalignment, we could get the test data which show the coefficient D_0 is relative to the SDFG's misalignment. Therefore, it's necessary to take the test error facter into consideration when selecting the model equation. Meanwhile, it is useful to pay enough attention to reduce misalignment during design, manufacting and testing. In the end of the paper, the coefficient was analysed by statistical test, and the results as reference verified that the coefficient D_0 is in creditable.
出处 《导航与控制》 2017年第5期52-57,共6页 Navigation and Control
关键词 三浮陀螺仪 一次项误差系数 显著性分析 安装误差角 single degree of freedom rate integrating gyro(SDFG) coefficient attributable to acceleration prominence analysis misalignment
  • 相关文献

参考文献8

  • 1梅硕基..惯性仪器测试与数据分析[M].西安:西北工业大学出版社,1991:248.
  • 2胡恒章编..陀螺仪漂移测试原理及其实验技术[M].北京:国防工业出版社,1981:152.
  • 3于波等编著..惯性技术[M].北京:北京航空航天大学出版社,1994:311.
  • 4陈桂明著..液浮陀螺仪性能改进技术[M].北京:科学出版社,2014:224.
  • 5吴广玉等编著..系统辨识与自适应控制 下[M].哈尔滨:哈尔滨工业大学出版社,1987:199.
  • 6何晓群..应用回归分析 第4版[M].北京:中国人民大学出版社,2015.
  • 7刘建波,魏宗康.石英加速度计误差系数显著性分析[J].中国惯性技术学报,2011,19(5):615-620. 被引量:8
  • 8严恭敏编..惯性仪器测试与数据分析[M].北京:国防工业出版社,2012:265.

二级参考文献9

共引文献7

同被引文献6

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部