摘要
介绍了皂膜流量计校准过程中,以干式流量校准器为标准器时的湿度修正方法。通过对校准系统的理论分析和试验验证,得出了有效可靠的修正系数,可以消除湿度对校准结果的影响。
The method to calibrate the humidity in the calibration of soap flm fow meter when the standard device is dry fow calibrator is given. Theoretical analysis and test verifcation have been done in the calibration system. The effective and reliable correction coeffcient is obtained. Then the effect of humidity on calibration results can be eliminated.
作者
朱碧玉
陈超
张进明
Zhu Biyü;Chen Chao;Zhang Jinming(Shanghai Institute of Measurement and Testing Technolog)
出处
《上海计量测试》
2018年第2期43-44,47,共3页
Shanghai Measurement and Testing
关键词
皂膜流量计
湿度
修正
soap flm fow meter
humidity
correction