常规超声结合弹性成像面积比值法在乳腺BI-RADS 4类病变诊断中的应用 被引量:10

Conventional ultrasound combined with elastography area ratio in diagnosing breast BI-RADS 4
下载PDF
导出
摘要 目的探讨弹性成像面积比值在乳腺BI-RADS 4类病变中的应用价值。方法选取96例患者115个BI-RADS 4类乳腺肿块行病变最大横径切面(A面)、最大前后径切面(B面)、最大上下径切面(C面),采用面积比值法分析弹性成像图像,判断不同切面在乳腺BI-RADS 4类病变中的实际价值。结果在BI-RADS 4类病变良性、恶性肿块面积比值差异有统计学意义(P<0.05);A、B、C切面面积比值良恶性肿块判断临界点分别为1.32、1.26与1.53。结论在乳腺BI-RADS 4类病变肿块良恶性判断中,不同切面弹性成像面积比值法的应用价值高。 Objective To study the elasticity imaging in breast area ratio value of BI-RADS4 type lesions.Methods 96 patients,115 BI-RADS 4 breast lumps line lesion( A surface) on the maximum transverse diameter section,after the largest diameter section( B),( C),on the top and bottom diameter section of the biggest area ratio method was used to analyze the elastic imaging image,determine different aspects in breast BI-RADS4 type of pathological changes in the actual value. Results In the BI-RADS4 class lesions,benign and malignant mass area ratio was significant difference( P < 0. 05);A,B,C surface area ratio of benign and malignant tumors determine critical point were 1. 32,1. 26 and 1. 53 respectively. Conclusions The value of different area ratio method of plane elasticity imaging is high in BI-RADS4 type lesions.
出处 《中国临床保健杂志》 CAS 2018年第1期104-106,共3页 Chinese Journal of Clinical Healthcare
关键词 乳腺肿瘤 弹性成像技术 超声检查 多普勒 彩色 Breast neoplasms Elasticity imaging techniques Ultrasonography doppler color