摆位误差对体重减轻鼻咽癌患者脑干剂量影响 被引量:4

Effects of setup error on brainstem dose distribution in patients with nasopharyngeal carcinoma and weight loss
原文传递
导出
摘要 目的 探讨CBCT获得的摆位误差对脑干剂量影响.方法 局部晚期鼻咽癌患者4例行3DRT并获取在线摆位误差,计划33次放疗,脑干最高总量限量〈54 Gy,脑干单次最高限量为163.6 cGy.将原计划中心点根据摆位误差进行偏移,带入原计划重新运算.每位患者单次治疗后记录脑干每一层面〉163.6 cGy区域.再通过CORELDRAW软件将同一脑干层面超剂量区域进行叠加,用Photoshop软件统计超剂量区域.记录患者放疗前体重变化.配对样本t检验差异.结果 未经校正的116次CBCT在x、y、z方向和床角上的误差分别为(0.90±0.85)、(0.98±0.6)、(0.98±1.05)mm和0.25°±0.49°.116次CBCT共获取2502层脑干图像,其中2040层脑干剂量不同程度超过最高限量;其中1例患者脑干超剂量次数最多,20次以上的总体积达3.83%;1例患者脑干超剂量体积仅占总体积的0.13%.体重减少和摆位误差之间无相关性(P〉0.05).结论 CBCT获得的摆位误差对脑干剂量有明显影响,它是一种有效的验证手段,对勾画靶区具有指导意义. Objective To investigate the effects of setup error acquired by cone-beam computed tomography ( CBCT) on the dose to the brainstem. Methods Setup errors were collected from 4 patients with locally advanced nasopharyngeal carcinoma ( NPC) undergoing three-dimensional radiotherapy. In the 33 radiotherapy plans, the upper limit of the dose to the brainstem was 54 Gy in total radiation and 163. 3 cGy in a single radiation. Treatment was re-planed using a corrected center based on setup error. For each patient, areas exposed to radiation higher than 163. 64 cGy were recorded in each layer of the brainstem after single treatment. The CorelDraw software was used to superpose all overdose areas in each layer of the brainstem. The Photoshop software was used to calculate the volume of the overdose areas. The changes in weight after radiotherapy were recorded and analyzed using paired t test. Results In 116 CBCT scans without correction, the setup errors were (0.90±0. 85),(0.98±0. 6), and (0.98±1. 05) mm in x, y, and z directions, and (0.25±0. 49)° in rotation. According to the images of a total of 2502 layers of the brainstem acquired by 116 CBCT scans, 2040 layers of the brainstem had doses over the upper limit to different degrees. The patient with the highest frequency of overdose in the brainstem had a total overdose volume up to 3. 83% in 20 scans. Another patient had the overdose volume only 0. 13% of the total volume. There was no significant relationship between weight loss and setup error ( P〉0. 05 ) . Conclusions Setup error acquired by CBCT has a substantial impact on the dose to the brainstem. It is an effective approach for verification of target volume delineation.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2016年第12期1331-1334,共4页 Chinese Journal of Radiation Oncology
基金 湖南省科技计划项目(2011SK3220)
关键词 体层摄影术 X线 锥形束 脑干剂量 摆位误差 鼻咽肿瘤/三维放射疗法 Tomography X-ray cone-beam Brain stem dose Set-up Nasopharyngeal neoplasms/three-dimensional radiotherapy
  • 相关文献

参考文献1

二级参考文献14

  • 1Xia P, Fu KK, Wong GW, et al. Comparison of treatment plans involving intensity-modulated radiotherapy for nasopharyngeal carcinoma. IntJ Radiat Oncol Biol Phys,2000,48(2) :329-337. 被引量:1
  • 2Wu Q, Manning M, Schmitt-Ullrich R, et al. The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head and neck cancers: a treatment design study. Int J Radiat Oncol Biol Phys,2000;46(1) :195-205. 被引量:1
  • 3Kam MK, Chau RM, Suen J, et al. Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetrie advantage over conventional plans and feasibility of dose escalation. Int J Radiat Oncol Biol Phys, 2003 ,56 (1): 145-157. 被引量:1
  • 4Saarilahti K, Kouri M, Collan J, et al. Intensity modulated radiotherapy for head and neck cancer: evidence for preserved salivary gland function. Radiother Oncol, 2005; 74 (3): 251- 258. 被引量:1
  • 5Barker JL Jr, Garden AS, Ang KK, et al. Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/ linear accelerator system. Int J Radiat Oncol Biol Phys,2004; 59(4) :960-970. 被引量:1
  • 6Hansen EK, Bueci MK, Quivey JM, etal. Repeat CT imaging and replanning during the course of IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys,2006;64(2) :355-362. 被引量:1
  • 7Oldham M, Letourneau D, Watt L, et al. Cone-beam-CT guided radiation therapy: a model for on-line application. Radiother Oncol, 2005 ; 75 (3) : 271-278. 被引量:1
  • 8Sharpe MB, Moseley DJ, Purdie TG, et al. The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator. Med Phys,2006;33(1) :136-144. 被引量:1
  • 9Ding GX, Duggan DM,Coffey CW, et al. A study on adaptive IMRT treatment planning using kV cone-beam CT. Radiother Oneol,2007;85(1) :116-125. 被引量:1
  • 10Lee C, Langen KM, Lu W, et al. Evaluation of geometric changes of parotid glands during head and neck cancer radiotherapy using daily MVCT and automatic deformable registration. Radiother Oncol,2008,89(1):81-88. 被引量:1

共引文献7

同被引文献46

引证文献4

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部