摘要 目的了解全省医用电子加速器输出量的准确性,促进放射治疗质量和安全。方法依据相关标准与方法对全省部分医用电子加速器的输出量进行检测。结果医用电子加速器输出量初检合格率88.2%,复检后全部合格,合格率100%;进口设备输出量检测结果明显优于国产设备,三级医院和省级医院的合格率明显高于二级医院和市、县级医院。结论检测结果表明河北省医用电子加速器的输出量的准确性还有待提高,应进一步采取措施加强放射治疗质量控制工作。 Objective To investigate the accuracy of the province's output of medical electron accelerators and promote the quality and safety of radiotherapy.Methods According to the relevant standards and methods,some of the province's output of medical electron accelerators were tested.Results The passing rate of the medical electron accelerator output of the first test was of 88.2%.All qualified after re-examination,with a passing rate of 100%.Imported equipment output test results significantly better than that of the domestic equipment.The passing rate of 3 lever hospitals and the provincial hospitals was significantly higher than the 2-level hospitals and city-level,county-level hospitals.Conclusion Test results show that the accuracy of the output of Hebei Medical Electron Accelerator should be further be improved.Further measures should be taken to strengthen the quality control of radiotherapy.
作者 张京战 周开建 高艳辉 郭大伟 ZHANG Jing - zhan ZHOU Kai - jian GAO Yan- hui GUO Da - wei(Hebei Province Center for Disease and Prevention, Shijiazhuang 050021 China)
出处 《中国辐射卫生》 2016年第5期567-568,572,共3页 Chinese Journal of Radiological Health
关键词 医用电子加速器 输出量 质量控制 Medical Electron Accelerator Radiation Output Quality Control
  • 相关文献

参考文献5

二级参考文献16

  • 1程金生,姜庆寰,郭朝晖,李开宝.^(60)Co放疗水平国家二级剂量标准的IAEA/WHO国际比对[J].中华放射医学与防护杂志,2004,24(4):370-371. 被引量:4
  • 2程晓军,张钦富,戴富友,乔红兵,胡传朋,卢国甫.河南省放射治疗防护现状调查与分析[J].中国辐射卫生,2007,16(1):61-62. 被引量:4
  • 3GB15213-1994,医用电子加速器性能和实验方法[s]. 被引量:4
  • 4JJG 589-2008,医用电子加速器辐射源[S]. 被引量:2
  • 5ICRU Report No. 24, Determination of absorbed dose in a pa- tient irradiated by beams of X - or γ - rays in radiotherapy procedures[ R]. 1976. 被引量:1
  • 6JJG589-2001.外照射治疗辐射源[S].[S].,.. 被引量:4
  • 7IAEA TRS No. 277. Absorbed dose determination in photon and electron beams. An International Code of Practice. Vienna: IAEA, 1987. 被引量:1
  • 8IAEA TRS No. 277 (Second Edition). Absorbed dose determination in photon and electron beams. An International Code of Practice. Vienna:IAEA, 1997. 被引量:1
  • 9ICRU Report 23. Measurement of absorbed dose in a phantom irradiated by a single beam of X or gamma Rays. ICRU, 1973. 被引量:1
  • 10国家计量局.卫生部&lt,关于肿瘤放射治疗剂量学的若干规定&gt,[Z].,1984.. 被引量:1